DOE-STD-1095-2018, Department of Energy Laboratory Accreditation Program for Personnel Dosimetry
This standard establishes the technical basis for the Performance Testing Laboratory (PTL) that administers the external dosimeter program for DOE site dosimetry programs seeking DOELAP accreditation. The performance testing categories for whole body dosimetry are based on ANSI/HPS N13.11, American National Standard for Dosimetry – Personnel Dosimetry Performance – Criteria for Testing, and for extremity dosimetry from ANSI/HPS N13.32, American National Standard for Performance Testing of Extremity Dosimeters.
Supersedes:
Superseded By:
Version history and related documents
Superseded by
A newer version replaces this document.
Supersedes
Earlier documents this one replaced.
Document text
Text extracted from the attached file. Refer to the original document for the authoritative version.
Section 1
ii
NOT MEASUREMENT
SENSITIVE
DOE-STD-1095-2018
October 2018
DOE STANDARD
DEPARTMENT OF ENERGY LABORATORY
ACCREDITATION PROGRAM FOR PERSONNEL
DOSIMETRY
U.S. Department of Energy AREA SAFT
Washington, DC. 20585
DISTRIBUTION STATEMENT A. Approved for public release; distribution is unlimited.
DOE-STD-1095-2018
ii
Available to the public on the DOE Technical Standards Program Web Site at
https://www.standards.doe.gov
DOE-STD-1095-2018
iii
FOREWORD
The Department of Energy (DOE) implemented the DOE Laboratory Accreditation Program (DOELAP) for
external dosimetry in 1986. The objectives of the DOELAP external dosimetry program is to assure the
competency of dosimetry measurements through calibration intercomparisons, performance testing,
on-site assessments; and to encourage applied research in areas where there is a technology shortfall.
DOE also expects the program to enhance cooperation and technical information exchange among its
sites and facilities to provide a more standardized and uniform radiation dosimetry capability. DOE sites
and facilities are expected to use standards and other technical guidance from the Department to
ensure that the performance of personnel dosimetry programs are adequate to meet the requirements
of Title 10, Code of Federal Regulations, Part 835, Occupational Radiation Protection and related
documents.
This standard establishes the technical basis for the Performance Testing Laboratory (PTL) that
administers the external dosimeter program for DOE site dosimetry programs seeking DOELAP
accreditation. The performance testing categories for whole body dosimetry are based on ANSI/HPS
N13.11, American National Standard for Dosimetry – Personnel Dosimetry Performance – Criteria for
Testing, and for extremity dosimetry from ANSI/HPS N13.32, American National Standard for
Performance Testing of Extremity Dosimeters.
Throughout this standard, the word “shall” is used to denote a required action that is to be performed,
and the word “should” is used to denote an action that is expected to be performed unless
documentation is provided validating technical equivalence.
DOE-STD-1095-2018
iv
Table of Contents
1 PURPOSE AND SCOPE .............................................................................................................1
2 APPLICABILITY .......................................................................................................................1
3 ACCREDITATION PROCESS ......................................................................................................1
3.1 Application ....................................................................................................................................... 1
3.2 Performance Testing ........................................................................................................................ 2
3.3 On-site Assessment .......................................................................................................................... 4
4 QUALITY ASSURANCE .............................................................................................................4
4.1 Quality Assurance Program ............................................................................................................. 4
4.2 Program Management ..................................................................................................................... 5
Section 2
4.3 Personnel Training and Qualifications ............................................................................................. 6
4.4 Documents and Records .................................................................................................................. 7
4.5 Work Processes ................................................................................................................................ 7
4.6 Quality Improvement ....................................................................................................................... 8
4.7 Facilities and Equipment .................................................................................................................. 8
4.8 Maintenance and Calibration ........................................................................................................ 11
4.9 Reporting ....................................................................................................................................... 12
APPENDIX A - REFERENCES ............................................................................................................. A-1
APPENDIX B – GUIDANCE FOR PROGRAMS THAT USE SERVICE PROVIDERS ...................................... B-1
DOE-STD-1095-2018
1
1 PURPOSE AND SCOPE
This technical standard describes the U.S. Department of Energy Laboratory Accreditation
Program (DOELAP) for external dosimetry, in support of worker health and safety. DOELAP
accreditation involves performance testing of dosimeters and the documentation of program
elements important to the long-term quality assurance of a dosimetry program and its ability
to accurately measure, record, and report occupational whole body and extremity dose.
Performance testing of dosimeters is conducted in a laboratory setting by the PTL. DOELAP
does not evaluate the adequacy of a dosimetry program to accurately measure occupational
dose in actual work environments encountered at DOE sites. The information in this technical
standard is intended for use by accredited programs, programs seeking accreditation, the
Performance Testing Laboratory, DOELAP Assessors, vendors, and subcontractors for
implementation of the DOELAP requirements, including the technical and quality assurance
aspects of the accredited program.
2 APPLICABILITY
This technical standard applies to DOE Headquarters, field organizations, and contractors
working to the individual monitoring requirements of 10 CFR Part 835. For the purposes of
accreditation, external dosimetry includes both whole body and extremity dosimeters that are
routinely used in the field to measure and determine a dose of record; It does not apply to
non-routine dosimetry.
3 ACCREDITATION PROCESS
To be granted accreditation, the following is required:
• Timely submittal of an application in accordance with the timeline set by the PTL;
• Compliance with DOELAP requirements contained in this standard;
• Demonstration of proficiency in processing each dosimeter model and type applied
for; and
• Successful passing of an on-site assessment conducted by qualified DOELAP Assessors.
3.1 Application
(a) Consideration for DOELAP accreditation requires the submission of an application,
including the documented quality assurance program. Initial applicants shall also submit a
program Self-Assessment.
(b) The program self-assessment is a documented internal review conducted by the applicant,
Section 3
which compares the program’s compliance status to the requirements set forth in this
DOE-STD-1095-2018
2
standard. Lines of inquiry for the self-assessment can be obtained through the DOELAP
Senior Technical Manager.
3.2 Performance Testing
(a) Proficiency in processing shall be demonstrated for each model and type of dosimeter
that the program intends to use to demonstrate compliance with 10 CFR § 835.402. The
radiation categories selected in the application for which accreditation is desired shall be
representative of the radiation type and energy encountered at any location where the
dosimeter will be used.
(b) Performance testing compares the results from the applicant’s reported measurement
values to the dose delivered to the applicant’s dosimeters using the criteria established in
the ANSI/HPS N13.11 and ANSI/HPS N13.32 standards. Test participants submit 15
dosimeters (5 per round) for each subcategory selected for evaluation, except for the
subcategories of category II whole body, which requires 21 dosimeters (7 per round).
(c) The delivered doses are not revealed to the applicants until after their results are reported
for all three rounds of irradiations. Comparing the reported dosimeter results from the
applicant’s measurement process with the delivered doses provides evidence of the
program’s proficiency in terms of both accuracy and precision. A variation beyond
established acceptance criteria provides a means for denying accreditation or granting
only accreditation for use of categories and dosimeters that successfully passed
performance testing.
(d) A minimum of 13 dosimeters shall be reported for each subcategory, except for the
subcategories of category II whole body, which requires a minimum of 18 dosimeters to
be reported. If the required minimum is not met, then analysis will be delayed until
replacement dosimeters undergo performance testing.
(e) The applicant shall review the performance testing data for potential improvements in the
dosimetry measurement system.
(f) Processing of performance testing dosimeters shall be defined and consistent with routine
processing procedures. The same dosimeter model, type, and sensitive element used to
assess occupational exposures shall also be used during performance testing.
(g) Performance Testing of additional dosimeter models and types not currently used in the
program may be requested through the Performance Testing Laboratory.
DOE-STD-1095-2018
3
3.2.1 Whole Body Dosimeter Performance Testing
(a) Performance testing of whole body dosimeters shall be conducted in accordance with
ANSI/HPS N13.11.
(b) Retesting is required if the performance testing results for any selected category from the Test
Category column do not meet criteria. The retest sequence is listed in Table 3-1. If a dose
algorithm was modified in response to a category failure, then retesting in all applied-for
categories is required.
Table 3-1. Retesting Requirements for Whole Body Dosimeter Performance Testing
Test Category Required Retesting
I. Accidents, photons I. Accident, Photons
II. Photons/photon mixtures
II. Photons/photon mixtures
II. Photons/photon mixtures
III. Betas
IV. Photon/beta mixtures
III. Betas II. Photons/photon mixtures
III. Betas
IV. Photon/beta mixtures
IV. Photon/beta mixtures II. Photons/photon mixtures
III. Betas
IV. Photon/beta mixtures
V. Neutron/photon mixtures II. Photon/photon mixtures
V. Neutron/photon mixtures
Section 4
(c) An applicant is allowed a maximum of 2 retests, irrespective of which whole body
performance testing category failed. Failure of the second retest will result in failure of
the application for whole body dosimetry accreditation.
3.2.2 Extremity Dosimeter Performance Testing
(a) Performance testing of extremity dosimeters shall be conducted in accordance with
ANSI/HPS N13.32.
(b) Retesting is required if the performance testing results for any selected category in the
Test Category column do not meet criteria. The request sequence is listed in Table 3-2. If
a dose algorithm was modified in response to a category failure, then retesting in all
applied-for categories is required.
DOE-STD-1095-2018
4
Table 3-2. Retesting Requirements for Extremity Dosimeter Performance Testing
Test Category Required Retesting
I. High-dose, photons I. High-dose, photons
II. Photons (IIA, IIB, or IIC)
II. Photons
II. Photons
III. Betas
IV. Beta/photon mixtures
III. Betas II. Photons
III. Betas
IV. Beta/photon mixture
IV. Beta/photon mixtures II. Photons
III. Betas
IV. Beta/photon mixture
(c) An applicant is allowed a maximum of 2 retests, irrespective of which extremity
performance testing category failed. Failure of the second retest will result in failure of
the application for extremity dosimetry accreditation.
3.3 On-site Assessment
An on-site assessment of an applicant’s program is conducted initially and triennially
thereafter to ensure a program meets the quality assurance requirements prescribed in this
standard. For initial accreditation, an on-site assessment is conducted after performance
testing is completed. A monitoring visit may also be conducted after implementation of a new
program or if major deficiencies were identified during an on-site assessment of an
established program.
4 QUALITY ASSURANCE
4.1 Quality Assurance Program
(a) The program shall have a documented quality assurance program describing the internal
management structure, system of procedures, and practices to ensure dosimetry results
are accurate, repeatable, verifiable, and properly recorded.
(b) The program’s quality assurance manual or supporting documentation shall include
• statement of quality policy and quality objectives;
• documented processes, procedures, and instructions;
• documents needed to ensure effective planning, operation, and control of processes;
• records required to demonstrate compliance with the quality assurance program;
DOE-STD-1095-2018
5
• dosimetry specifications and Technical Basis Documentation;
• acceptance criteria for dosimeter materials and holders;
• training objectives and processes for maintaining proficiency; and
• practices for handling and resolving contested dosimetry data and test reports.
4.2 Program Management
(a) Managerial and technical personnel shall have the resources needed to carry out their
duties, including the implementation of the Quality Assurance Program.
(b) A technical lead (however named), who is experienced in applied radiation dosimetry and
knowledgeable in the design and operation of the dosimetry system(s) currently used,
shall be assigned. The technical lead is responsible for ensuring that dosimetry data are
approved and validated. This includes making decisions regarding questionable data.
(c) A quality assurance (QA) lead (however named), who has responsibility and authority for
Section 5
ensuring that the quality assurance program is implemented, shall be assigned. The QA
lead shall have authority to communicate quality assurance issues directly with the
technical lead and other organizational management. The program technical lead may
function as the QA lead as long as the responsibilities are clearly defined.
(d) Responsibilities for the implementation of the quality assurance program shall be defined,
including the organizational structure and functional responsibilities of key personnel.
(e) The individuals responsible for the implementation of the quality assurance program may
delegate work to others but shall retain responsibility.
(f) Management and personnel shall be free from undue internal and external influences that
may adversely impact the quality of their work.
(g) Management shall conduct a formal review of the External Dosimetry Quality Assurance
Program at the midpoint (between year 1 and 2) of the DOELAP assessment cycle.
Management shall consist of the QA lead, the technical lead, and a member of senior
management that has authority for allocation of resources. The review shall include
assessing opportunities for improvement and the need for changes to policies or
processes. At minimum, the review shall include
• comparison of quality objectives and standards against achievements;
• assessment and test results;
• non-conformances and corresponding corrective actions, preventative measures, and
DOE-STD-1095-2018
6
deficiency trends;
• results from external and internal audits; and
• other relevant factors, such as quality control activities, resources, and training.
(h) A program shall have a documented plan for continuity of operations. This includes
service contracts, in-house maintenance, spare parts capabilities, and unexpected loss of
key personnel.
(i) When more than one organization is involved in the implementation of the requirements
for DOELAP accreditation (e.g., major equipment maintenance, calibration, document
control and records); the responsibilities, interfaces, and authority of each organization
shall be clearly defined and documented.
(j) When a vendor or subcontractor is involved in the implementation of the requirements
for DOELAP accreditation, the accredited program shall have a procedure describing how
they will ensure that all of the DOELAP requirements are maintained.
(k) External audits of a vendor or subcontractor’s quality assurance plan shall be performed
initially and at least once during the DOELAP accreditation period. Audits should be
performed at least one year prior to the DOELAP on-site assessment to allow assessors
adequate time to evaluate the program’s progress in managing corrective actions and final
resolutions of identified issues. The audits shall be supplemented by an ongoing
evaluation of the performance of the vendor or subcontractor through blind audits, which
are outlined in section 4.7.2.
4.3 Personnel Training and Qualifications
(a) All personnel performing accredited activities shall have the training, qualifications, and
competence to perform their assigned tasks effectively.
(b) A training program commensurate with the complexity and scope of the assigned
responsibilities shall be documented. Training shall be provided to achieve initial
proficiency, maintain proficiency, and adapt to changes in job responsibilities, new
technologies, or policies and procedures.
(c) The technical lead shall initially and at least annually evaluate and document the
Section 6
proficiency of each staff member authorized to perform dosimetry related functions. This
proficiency assessment shall include an observation of performance.
(d) In the event that proficiency is not achieved or maintained, any person’s work duties that
impact the quality of accredited activities shall be performed under the direction or
supervision of a properly trained and qualified individual. Such personnel shall not be the
DOE-STD-1095-2018
7
primary signatory on dose processing records or QA/QC reports until proficiency is
demonstrated.
4.4 Documents and Records
(a) A system shall be in place which clearly describes the process applied for controlling the
dosimetry documents and records throughout the entire dosimetry cycle.
(b) All documents that form the quality assurance program shall be controlled to ensure that
the correct and most current documents are being employed. Documents shall be
reviewed for accuracy and approved by authorized personnel in accordance with
documented internal review frequencies.
(c) A comprehensive record of processing activities shall be maintained. Records shall
contain sufficient identification to allow correlation with calibration and quality control
records.
(d) Procedures shall be established and maintained for the identification, collection, indexing,
access, filing, storage, maintenance, and disposal of quality and technical records.
(e) All quality assurance and technical records shall be legible, easily retrievable, and stored in
a suitable environment to prevent damage, deterioration, or loss. Records shall be
available for review during the on-site assessment.
(f) Electronic records shall be protected and regularly backed-up on a pre-determined
schedule to prevent unauthorized access, amendment, or loss.
4.5 Work Processes
(a) All accredited activities that can influence the assignment of dose to an individual shall be
conducted in accordance with established procedures, which shall include the following:
• work methods and sequence;
• equipment to be used;
• work environment;
• quality control;
• acceptance criteria;
• inspection points; and
• recordkeeping.
(b) Work process procedures shall control the preservation of identification of dosimeters,
DOE-STD-1095-2018
8
measurements, dose records, and other data on which the dose is based, and maintain
their traceability to the individual concerned.
(c) Work process procedures shall prescribe specifications and precautions to control the
processing, handling, issuing, storage, retrieval, and shipment of dosimeters.
4.6 Quality Improvement
(a) Quality control procedures shall be implemented to ensure that the equipment performs
at the levels of precision and accuracy defined in the processing protocols. Quality control
data shall be recorded in such a way that trends are detectable.
(b) When quality control data is found to be outside pre-defined acceptance criteria,
corrective actions to correct the problem and to prevent incorrect results from being
reported shall be documented. Reevaluation of all dosimeters processed since last
acceptance shall be performed.
(c) Software verification and validation (V and V) shall be performed in accordance with an
appropriate, documented software quality assurance plan. V and V shall include process
control software, dose algorithms, data processing, and record keeping. In addition,
software version control shall be included in the program’s documented control
procedures.
Section 7
(d) When computer or laboratory information systems are used to input, store, calculate, or
retrieve data in relation to key dosimeter processing steps, the program shall
• establish and maintain procedures describing the processes;
• validate the accuracy of data entry; and
• verify the accuracy of any calculations performed.
(e) The variability of test results among staff, equipment, and locations shall be assessed to
ensure consistency.
(f) Internal audits shall be conducted at least annually and structured in a way to ensure that
all elements of this standard are reviewed over the three year accreditation period. All
audits and actions taken for correcting identified problems and preventative actions
implemented to prevent recurrence shall be documented.
4.7 Facilities and Equipment
(a) Facilities and equipment shall be adequate to perform the type(s) of processing for which
DOE-STD-1095-2018
9
accreditation is sought. A list and description of facilities and equipment which have the
potential to impact the quality of dose results shall be available for review.
(b) Adequate facilities and equipment shall have the following:
• sufficient space to perform processing;
• proper shielding of areas from unwanted radiation;
• environmental monitoring and controls, including background radiation; and
• properly calibrated equipment.
(c) Adequate backup equipment shall be possessed and maintained in the event the primary
systems fail. If backup equipment or systems are not available, the program shall have
documented provisions to utilize the services of another DOELAP accredited laboratory in
an emergency.
4.7.1 Dosimeters
(a) A design specification shall be established for each dosimeter model and configuration.
The specification shall include dosimeter holders, any filter material used, the areal
density (mg/cm2) of the material, and the positions of the dosimetric material within the
dosimeter.
(b) Dosimeter materials and holders shall be acceptance tested before being placed into
service.
(c) The impacts of the following system characteristics shall be determined. Documentation
shall clearly indicate algorithm name and version used to generate the results.
• lower limit of detection;
• useful dose range;
• background contribution to dose equivalent;
• processing system measurement uncertainty;
• repeatability/precision;
• residual signal;
• angular dependence; and
• batch homogeneity.
(d) Fading of dosimeter materials under normal conditions shall be determined for two times
the period of intended use, not to exceed 6 months past the period of intended use. For
example, fading of quarterly dosimeters shall be documented and accounted for over the
period of 6 months.
DOE-STD-1095-2018
10
(e) Dosimeters placed into service shall be checked according to a defined schedule or
frequency to ensure all necessary components are in place. A screening procedure shall
be used to ensure dosimetry materials, including sensitive elements, are consistent with
the dosimeter design. Procedures shall include the phosphor type and sensitivity.
(f) Loading of dosimeters shall be carried out in a well-defined order to ensure the dosimeter
is in compliance with the design specification and prevent confusion in handling visually
similar elements. Precautions shall be taken to avoid optical fading and non-radioactive
contamination of the phosphor or the detector.
(g) If a dosimeter is used in radiation fields it is not designed for (e.g., a photon dosimeter
Section 8
being used in a mixed photon/neutron field) the effect of the radiation not intended to be
measured shall be determined.
4.7.2 Processing
(a) A positive system for identifying and tracking all dosimeters through the processing cycle
shall be established.
(b) Dosimeter reader operation and stability shall be verified before use with quality control
dosimeters and measurement of system internal parameters (e.g., photomultiplier tube
sensitivity, dark counts, light source counts). Records shall indicate that dose
measurements are made only with stable equipment.
(c) Annealing of dosimeters shall be conducted in a reproducible manner regarding time,
temperature, cooling rate, humidity, and light. For TLDs, it is preferred that thermal
erasing procedures be carried out in ovens reserved strictly for dosimeter annealing;
however, in-reader annealing can be done when very low irradiation doses have been
measured and when the in-reader annealing has been demonstrated to be reproducible.
The annealing technical basis shall be documented to demonstrate the upper dose range
limit for which annealing may be performed.
(d) Quality Control and unirradiated dosimeters, shall be used to routinely identify reader
processing problems. Each processing protocol shall provide for interspersing quality
control dosimeters. Records shall demonstrate reproducibility for the irradiation method.
Unirradiated and quality control dosimeter use frequency shall be determined based upon
the total number of dosimeters processed, equipment stability, type of quality control
checks, or other suitable method.
(e) Blind testing shall be conducted to validate the overall performance of the dosimetry
system. The blind testing program shall consist of the use of dosimeters irradiated by NIST
DOE-STD-1095-2018
11
traceable isotopic sources or x-ray beams to doses that are unknown to the processor.
Exposures to blind test dosimeters shall include those sources and x-ray beams for which
the program is accredited. Procedures describing steps to be taken in the event that blind
testing results are outside of pre-established criteria shall be documented. Blind test
dosimeters should be incorporated into every dosimeter exchange.
(f) The dosimetry algorithm shall be documented in sufficient detail to indicate its validity for
dose interpretation. Documentation shall indicate algorithm name and version, and
include
• Fundamental data for creating and testing;
• Uncertainty analysis of the algorithm;
• Process controls used for algorithm development; and
• Attributes and limitations of the algorithm.
(g) Deviations from processing procedures, equipment or facilities shall be verified to ensure
no degradation of performance has occurred.
4.7.3 Interim Processing
Although interim processing of Optically Stimulated Luminescence (OSL) dosimeters is not for
the dose of record, decisions are made based on interim results which may impact the overall
dose to the worker. For interim processing, the following is required:
• Technical basis document determining signal depletion as a function of the number of
times the dosimeter is processed.
• Calibration of processing equipment shall not be less restrictive than the
manufacturer’s prescribed requirements.
• All personnel performing interim processing activities shall meet the requirements of
section 4.3. Personnel Training and Qualifications.
4.8 Maintenance and Calibration
Section 9
(a) A preventative maintenance program for equipment used to process dosimeters or
perform quality control checks shall be implemented.
(b) Equipment used for dosimeter processing or quality control shall be calibrated periodically
or whenever the accuracy of the equipment is suspect. Calibration procedures shall
identify required accuracy and define the methods and frequency for checking accuracy.
Calibration procedures shall not be less restrictive than the manufacturer’s prescribed
requirements. A technical basis shall be developed when calibration techniques differ
DOE-STD-1095-2018
12
from manufacturer recommendations or when calibration frequency is not prescribed by
the manufacturer.
(c) Processing-equipment calibration or verification records shall include
• Equipment name or description;
• Model, style, and serial number;
• Manufacturer;
• Notation of all equipment variables requiring calibration or verification;
• The range of the calibration or verification;
• The resolution of the instrument and its allowable error;
• Calibration or verification date and schedule;
• Date and result of last calibration;
• Identity of the laboratory individual or external service responsible for calibration;
• Source of reference standard and traceability; and
• Environmental conditions.
(d) Equipment shall be properly identified to correlate with calibration records and
maintenance logs.
(e) The energy response of each type or model of dosimeter shall be characterized for all
radiation categories and exposure ranges for which it is to be used.
(f) All calibrations and characterizations shall be performed using reference standards
traceable to the National Institute of Standards and Technology (NIST) national standards
or standards maintained by an equivalent national standards authority.
(g) All processing equipment calibration, verification, and maintenance practices shall be
documented.
(h) When results are found to be inaccurate, reviews of the equipment used to generate the
results shall be conducted to determine the validity of the data and the corrective actions
to be taken.
4.9 Reporting
The dose report (initial report from the dosimetry processor or other records) shall include
• Processor name and address if different from accredited program;
• Name of accredited program;
• Pertinent dates for the wear period and the identification of dosimeters;
DOE-STD-1095-2018
13
• Processor and accredited program identification codes, as appropriate;
• An explanation of any deviation from routine processing procedures if the deviation
could affect the reported dose;
• The signature of or reference to the technical lead (however named); and
• Software version(s) of the dose algorithm(s) used.
DOE-STD-1095-2018
A-1
APPENDIX A - REFERENCES
The current versions of the following documents allow for complete implementation of this technical
standard.
American National Standards Institute (ANSI). 2009. Personnel Dosimetry Performance – Criteria for
Testing. ANSI/HPS N13.11-2009. New York, NY.
American National Standards Institute (ANSI). 2008. Performance Testing of Extremity Dosimeters.
ANSI/HPS N13.32-2008. New York, NY.
U.S. Department of Energy. 2008a. Radiation Protection Programs Guide for use with Title 10, Code of
Federal Regulations, Part 835, Occupational Radiation Protection, DOE G 441.1-1C, Admin Chg 1.
Washington, D. C.
U.S. Department of Energy. 2011. Department of Energy Laboratory Accreditation Program for Personnel
Section 10
Dosimetry. DOE-STD-1095-2011. Washington, DC.
U.S. Department of Energy. 2011. Title 10, Code of Federal Regulations, Part 830, Nuclear Safety
Management. Washington, DC.
U.S. Department of Energy. 2011. Title 10, Code of Federal Regulations, Part 835, Occupational
Radiation Protection. Washington, DC.
DOE-STD-1095-2018
B-1
APPENDIX B – GUIDANCE FOR PROGRAMS THAT USE SERVICE PROVIDERS
DOELAP accredited programs may purchase dosimetry services from service providers; however, the
DOELAP accredited program has the responsibility for ensuring the requirements of this technical
standard are met. The purpose of this appendix is to outline the major considerations of a program that
is purchasing dosimetry services from a commercial dosimeter vendor or a DOE dosimeter processor.
A copy of the work agreement with the service provider, including any agreed upon commitments shall
be available for review. The work agreement should clearly establish
• Access to relevant documents, including dosimetry technical basis documents, policies and
procedures, and the documented quality assurance program;
• Personnel whole body and extremity dosimeters provided for beta and gamma radiation;
• Personnel whole body dosimeters for neutron radiation provided, including calibration that
closely represents the workplace neutron spectrum;
• External dosimetry data validation and verification;
• External dosimetry reports (see section 4.9);
• Emergency external dosimetry services; and
• Appropriate packaging and handling of dosimeters.
Staff shall have sufficient qualifications and experience to be able to
• Sufficiently assess the capabilities and limitations of the service provider;
• Validate dosimeter results used to determine dose-of-record;
• Provide oversight of the service provider including the review of quality control data and
conduct on-site assessments;
• Identify error trends and anomalous data; and
• Conduct quality assurance assessments.
A technical basis for the selected performance testing categories or subcategories shall be available.
The program shall have a procedure for conducting quality assurance assessments of the service
provider; including on-site audits, QC reviews, and blind audit dosimeters. The procedures shall also
describe how findings are identified and corrected.
The program shall have a procedure for handling and shipping dosimeters. The procedure shall include
details on maintaining dosimeter chain-of-custody and assessment of any transit dose.
1 PURPOSE AND SCOPE
2 APPLICABILITY
3 ACCREDITATION PROCESS
3.1 Application
3.2 Performance Testing
3.2.1 Whole Body Dosimeter Performance Testing
3.2.2 Extremity Dosimeter Performance Testing
3.3 On-site Assessment
4 QUALITY ASSURANCE
4.1 Quality Assurance Program
4.2 Program Management
4.3 Personnel Training and Qualifications
4.4 Documents and Records
4.5 Work Processes
4.6 Quality Improvement
4.7 Facilities and Equipment
4.7.1 Dosimeters
4.7.2 Processing
4.7.3 Interim Processing
4.8 Maintenance and Calibration
4.9 Reporting
APPENDIX A - REFERENCES
APPENDIX B – GUIDANCE FOR PROGRAMS THAT USE SERVICE PROVIDERS