DOE-STD-1054-96, Guideline to Good Practices for Control and Calibration of Measuring and Test Equipment (M&TE) at DOE Nuclear Facilities
Functional areas: Good Practices, Calibration, Measuring, Test Equipment, Nuclear Facilities
This guide is intended to assist facility maintenance operations in the review of existing and in developing new programs to ensure the accuracy and integrity of performance data derived from Plant process and control instrumentation is verified by controlled application of properly calibrated/certified measuring and test equipment (M&TE) having the
appropriate precision, design accuracy, and durability for their intended use.
Unknown Block text
Document text
Text extracted from the attached file. Refer to the original document for the authoritative version.
Section 1
NOT MEASUREMENT
SENSITIVE
DOE-STD-1054-93
March 1993
DOE STANDARD
GUIDELINE TO GOOD PRACTICES FOR
CONTROL AND CALIBRATION OF
MEASURING AND TEST EQUIPMENT
(M&TE) AT DOE NUCLEAR FACILITIES
U.S. Department of Energy AREA MNTY
Washington, D.C. 20585
DISTRIBUTION STATEMENT A. Approved for public release; distribution is unlimited.
WELCOME TO PDF
This Portable Document Format (PDF) file contains bookmarks, thumbnails, and hyperlinks to help you navigate through the document. All items listed on the Contents pages are linked to their corresponding sections. Click on the page numbers located at the bottom of each linked page to return to the Contents page. Click on the DOE Seal below to move to the Contents page.
This document has been reproduced directly from the best available copy.
Available to DOE and DOE contractors from the Office of Scientific and
Technical Information, P.O. Box 62, Oak Ridge, TN 37831.
Available to the public from the National Technical Information Service, U.S.
Department of Commerce, 5285 Port Royal Rd., Springfield, VA 22161.
Order No. DE93013952
DOE-STD-1054-93
FOREWORDFOREWORD
The purpose of the Guideline to Good Practices for Control and Calibration of
Measuring and Test Equipment (M&TE) at DOE Nuclear Facilities is to provide contractor
maintenance organizations with information that may be used for the development and
implementation of a rigorously controlled maintenance program directed at controlling and
calibrating M&TE used for maintenance tasks at DOE nuclear facilities. This document
is intended to be an example guideline for the implementation of DOE Order 4330.4A,
Maintenance Management Program, Chapter II, Element 11. DOE contractors should not
feel obligated to adopt all parts of this guide. Rather, they should use the information
contained herein as a guide for developing an M&TE program applicable to their facility.
iii
DOE-STD-1054-93
FOREWORD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iii
1. INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.1 Purpose . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.2 Background . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.3 Application . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
2. DEFINITIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2. 1 Acronyms Used in this Standard. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2. 2 Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2. 3 Certification. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2. 4 Gross Error. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2. 5 Measuring & Test Equipment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2. 6 Primary Standards. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2. 7 Recall Program. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2. 8 Secondary Standards. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2. 9 Standards Laboratory. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2.10 Systems Stations. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2.11 Working Standards. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Section 2
3. CONTROL AND CALIBRATION OF MEASURING
AND TEST EQUIPMENT (M&TE) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
3.1 Discussion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
3.2 Scope . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
3.3 Responsibilities. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
3.3.1 M&O Operators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
3.3.2 Plant Personnel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
3.4 Guidelines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
3.4.1 Procurement and Acceptance of M&TE . . . . . . . . . . . . . . . . . . . . . 7
3.4.2 M&TE Identification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
3.4.3 Control and Use of M&TE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
3.4.4 Calibration of M&TE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.4.5 M&TE Repairs/Out of Calibration . . . . . . . . . . . . . . . . . . . . . . . . 15
3.4.6 Records . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
APPENDIX A EXAMPLE OF "STICKERS" (TAGS) USED
IN M&TE CALIBRATION PROGRAM . . . . . . . . . . . . . . . . . . . A - 1
APPENDIX B MEASURING AND TEST EQUIPMENT
RECORD SHEET . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B - 1
v
DOE-STD-1054-93
APPENDIX C CALIBRATION INTERVAL CHANGE
AUTHORIZATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C - 1
APPENDIX D GROSS ERROR REPORT . . . . . . . . . . . . . . . . . . . . . . . . . . . . D - 1
APPENDIX E CONTROL AND CALIBRATION OF
MEASURING AND TEST EQUIPMENT
(M&TE) SAMPLE LESSON PLAN . . . . . . . . . . . . . . . . . . . . . . E - 1
vi
DOE-STD-1054-93
1. INTRODUCTION
1.1 Purpose
This guide is intended to assist facility maintenance operations in the review of
existing and in developing new programs to ensure the accuracy and integrity of performance
data derived from Plant process and control instrumentation is verified by controlled
application of properly calibrated/certified measuring and test equipment (M&TE) having the
appropriate precision, design accuracy, and durability for their intended use. It is expected
that each DOE facility may use different approaches or methods than those defined in this
guide. The specific guidelines that follow reflect generally accepted industry practices.
Therefore, deviation from any particular guideline would not, in itself, indicate a problem.
If substantive differences exist between the intent of the Guideline and actual practice,
management should evaluate current practice to determine the need to include/exclude
proposed features. A change to maintenance practice would be appropriate if a performance
weakness was determined to exist. Development, documentation, and implementation of
other features which further enhance these guidelines for specific applications, is encouraged.
Additional information pertinent to the implementation of this guideline is found in the
following DOE Guidelines:
1) "Guidelines to Good Practices for Procurement of Parts, Materials, and Services at
DOE Nuclear Facilities"
2) "Guidelines to Good Practices for Material Receipt, Inspection, Handling, Storage,
Retrieval, and Issuance at DOE Nuclear Facilities"
Section 3
Appendix E is provided for use by facility trainers who intend to provide training
regarding this element.
1.2 Background
The information in this guide was developed from commercial and DOE sources.
Each facility should select those details that are applicable, add any unlisted knowledge or
experience that are applicable, and develop and implement facility-specific M&TE
maintenance programs. Facilities that have existing documented M&TE maintenance
programs should review this guide to identify details that may enhance their existing
programs.
1
DOE-STD-1054-93
1.3 Application
The content of this guide is generally applicable to all DOE nuclear facilities.
Portions of the programs outlined may not be applicable to all facilities because maintenance
organizations, disciplines, titles, and responsibilities may vary among DOE nuclear facilities.
Facility maintenance personnel should verify the adequacy or improve existing maintenance
programs by adapting this guide to their specific facility and individual maintenance
disciplines.
2
DOE-STD-1054-93
2. DEFINITIONS
2. 1 Acronyms Used in this Standard. The acronyms used in this standard are defined as
follows:
a. M&TE - Measuring and Test Equipment
b. NIST - National Institute of Standards and Technology
c. SSC - Structures, Systems, and Components
2. 2 Calibration. The comparison of readings from the instruments being tested with
validated readings observed on the measurement standards.
2. 3 Certification. An indication by the appropriate authority that the deviations
determined in the calibration do not exceed specified limits.
2. 4 Gross Error. An out-of-tolerance condition for an instrument in the M&TE Program
which may result in an unacceptable product. When a plant guide has not been
prepared or does not list gross error limits, any condition outside the calibration
limits, defined by supervision in the group using the equipment, is considered a gross
error.
2. 5 Measuring & Test Equipment. M&TE includes all devices or systems used to
calibrate, certify, measure, gauge, troubleshoot, test, or inspect in order to control
data or to acquire data to verify conformance to specified requirements. M&TE does
not include permanently installed plant instrumentation, nor does it include test
equipment used for preliminary checks where data obtained is not used to determine
acceptability or verify conformance to established criteria.
2. 6 Primary Standards. Calibrated by the National Institute for Standards and
Technology (NIST) or other authoritative reference source. Its use should be
restricted to the standards laboratory.
2. 7 Recall Program. A system to recall and service measuring and test equipment
(M&TE).
2. 8 Secondary Standards. Calibrated by comparison with a primary standard of the same
measurement modes. It should be used by laboratory personnel and stored in the
standards laboratory. Use by other than standards laboratory personnel should be
limited to that approved by the applicable manager.
3
DOE-STD-1054-93
2. 9 Standards Laboratory. A standards laboratory is a central on-site facility that
maintains, calibrates, and certifies most of the plant portable instrumentation and test
equipment.
2.10 Systems Stations. A location where two or more instruments are used as a unit to
make certification measurements.
Section 4
2.11 Working Standards. Calibrated in the plant with a primary or secondary standard of
the same measurement mode, or calibrated using other measurement modes. It
generally should be used in day-to-day activities, (mainly in direct field applications
where direct/ready access is required or when ALARA considerations exist), to
certify product certification equipment and instruments in the maintenance recall
program using approved procedures.
4
DOE-STD-1054-93
3. CONTROL AND CALIBRATION OF MEASURING AND
TEST EQUIPMENT (M&TE)
3.1 Discussion
The control and calibration of measuring and test equipment (M&TE) plays an ever
increasing role in the safe operation of DOE Nuclear Facilities. The need to ensure proper
calibration and traceability of use has increased. Therefore, implementation of an M&TE
control and calibration program is essential. A computer based system should be considered
for the establishment of an M&TE program.
This guideline establishes a process to ensure that M&TE is properly controlled,
calibrated, and certified with standards and data traceable to NIST.
The M&TE program shall be comprehensive and ensure the following:
1) All M&TE is uniquely identified, calibrated, controlled, and provides accuracy
traceability.
2) A recall program maintains the total inventory and status of all M&TE.
3) Out-of-tolerance M&TE is removed from service.
4) Plant equipment calibrated with out-of-tolerance M&TE is evaluated in a timely
manner for impact on previous output, current operability and is re-calibrated as
necessary.
5) The critical parameters of all M&TE such as range, precision, accuracy, etc. are
documented; the documents shall be retrievable, and maintained as quality records.
6) Process repeatability.
M&TE intended only as an indicator during troubleshooting and not intended for
calibration or data measurement should not be required to display calibration/certification
identification.
The M&TE program should be regularly evaluated to:
a) Detect trends which adversely affect safe, reliable use of equipment
b) Determine opportunities for expanded applications of equipment
5
DOE-STD-1054-93
c) Determine new technology and its potential for adding value to applications if
acquired
3.2 Scope
This guideline applies to M&TE used on safety class items, as well as, safety-related
structures, systems, and components (SSC) or on SSC which affects critical plant
performance and reliability.
3.3 Responsibilities
3.3.1 M&O Operators
It is the responsibility of each M&O Operator to ensure that M&TE for their
group has a current calibration sticker prior to use.
3.3.2 Plant Personnel
Plant personnel using M&TE should be responsible for reporting any
malfunctions or suspected malfunctions to the cognizant department supervisor.
Personnel should ensure that the equipment calibration date has not expired. Refer
to Section 3.4.3.4.2 for appropriate action for equipment past the calibration due date.
6
DOE-STD-1054-93
3.4 Guidelines
3.4.1 Procurement and Acceptance of M&TE
3.4.1.1 Procurement
Copies of the technical manual with full schematics, troubleshooting
sections, spare parts lists, table of specifications, and calibration
procedures should be obtained with each type of equipment. A
certificate of calibration and tolerance should be included.
Distribution of these manuals shall be in accordance with document
control procedures. Training aids furnished by the manufacturer(s)
may be desirable for certain types of equipment, as determined by the
M&TE requestor. Equipment selected should be the proper range,
accuracy, and precision. Also, where practical, the equipment selected
should be a type that minimizes the possibility of human error (e.g.,
direct digit readout, large mirror scale, null indicators, and direct
temperature readout).
Section 5
3.4.1.2 Receipt Inspection and Acceptance Test
Prior to acceptance, a receipt inspection and test should be conducted
in accordance with DOE "Guidelines to Good Practices for
Procurement of Parts, Materials, and Services at DOE Nuclear
Facilities," "Guidelines to Good Practices for Material Receipt,
Inspection, Handling, Storage, Retrieval, and Issuance at DOE Nuclear
Facilities") and applicable plant policies and procedures.
3.4.2 M&TE Identification
3.4.2.1 Each department should maintain a master inventory list of the
applicable M&TE covered by this program. A standards laboratory
should maintain a plant level master list of applicable M&TE. These
listings should include the following information for each item:
a) Systems/Stations number - this identifies a specific document
which establishes parameters, range, precision, accuracy and
other requirements for application of the individual item listed.
b) Unique-Number
c) Date of last calibration
7
DOE-STD-1054-93
d) Expiration date
e) Applicable procedure number
f) Building/location
g) Cost collection account
h) Noun description
i) Owner organization
j) Calibration interval
3.4.2.2 For new equipment, the manager responsible for the equipment,
should assign an appropriate identification number and establish a
maintenance history file (see Section 3.4.6). The history file should
be maintained by the standards laboratory.
3.4.2.3 Each item of M&TE should be marked with a unique identification
number.
XXX XXX
Group Identification Unique Identification Number
IAC-Instruments and Control
MNT-Maintenance
Prefixes should be added to the unique identification number in order
to group equipment by type (e.g., G-XXX for gauges or TW-XXX for
torque wrenches). This designation should be etched into or
permanently attached to the equipment in a manner that should not
interfere with the operation or accuracy of the equipment. When this
designation cannot be attached to the equipment, it shall be attached
to the case with a means of identifying the equipment to the case
(e.g., with color coding).
8
DOE-STD-1054-93
3.4.3 Control and Use of M&TE
3.4.3.1 Tagging
3.4.3.1.1 A "calibration" sticker shall be attached to each M&TE
device. This sticker (see example Appendix A, Figure
1) should indicate: (1) date calibration performed
(mo/day/year); (2) badge number of the individual who
calibrated the device; and (3) expiration date
(mo/day/year). The expiration date is determined from
the interval and the date last calibrated. The
calibration sticker should be removed and replaced
with a new sticker each time the equipment is
calibrated.
3.4.3.1.2 A "restricted use" sticker should be placed on any
equipment requiring special restrictions or precautions
in its use. This sticker (see example Appendix A,
Figure 2) should be attached in addition to the
calibration sticker and should include (1) equipment
identification number, (2) a statement of the restrictions
that apply, (3) badge number of the individual
attaching the sticker, and expiration date, if applicable,
and (4) the date the sticker is attached.
3.4.3.1.3 A "rejected" sticker should be placed on any equipment
that fails to meet acceptance criteria or is suspected of
being defective. This sticker (see example Appendix
A, Figure 3) should include (1) device identification
number, (2) reason for rejection, (3) badge number of
the individual attaching the sticker. A "rejected"
sticker may also be used to identify devices that are
not to be used due to the device exceeding the
calibration due date.
Section 6
9
DOE-STD-1054-93
3.4.3.2 Storage
3.4.3.2.1 Department M&TE storage areas should be authorized
by the maintenance manager. All equipment should be
stored in a manner that assures integrity is maintained
and unintended contamination is minimized.
Equipment in these storage areas should be identified
as to its current status, per Section 3.4.3.1 of this
guideline. The M&TE storage areas should provide
sufficient separation of the ready-to-use equipment
(calibrated and restricted use) from other equipment
(expired) to preclude inadvertent use.
3.4.3.2.2 The environment of M&TE storage areas should be
controlled to preclude any adverse effect on equipment
accuracy. Environmental factors that should be
considered include, but are not limited to, temperature,
humidity, vibration, radio frequency interference,
electromagnetic interference, and fumes.
3.4.3.2.3 Security of the M&TE storage area should be
maintained by designated individuals responsible for
control of measuring and test equipment. In the
absence of these individuals, the storage area should be
locked with access controlled by the responsible
department manager and/or supervisor.
3.4.3.2.4 Instruments used on systems that contain oil, brackish
water, etc., should be cleaned thoroughly after each
use.
3.4.3.3 Restrictions on M&TE Use
3.4.3.3.1 The use of equipment containing mercury should be
prohibited on nuclear-related systems and should be
restricted to absolute necessity on all other systems.
3.4.3.3.2 In applications where a caustic environment is present,
the use of aluminum material in contact with the
process liquid should be prohibited.
3.4.3.3.3 Oil-free stainless steel components (e.g., tubing, valves,
Bourdon tubes, and pressure cells) should normally be
used in nuclear applications.
10
DOE-STD-1054-93
3.4.3.3.4 Instruments used on systems that contain oil, brackish
water, etc., should be identified clearly to avoid
contamination of other systems. For example an oil-
contaminated gauge should not be used on an oil-free
compressed air system, or a saltwater gauge should not
be used on a system containing process liquids.
3.4.3.4 Issue of M&TE for use
3.4.3.4.1 Issuing of M&TE should be conducted utilizing a
usage record. (See Appendix B.) The M&TE should
have a usage record filled out to identify each item
calibrated or tested by that M&TE. A new usage
record should be initiated each time the M&TE is
calibrated, and the previous record should be retained
in the maintenance history file. The usage record for
each item of M&TE should be retained in the issue
area.
3.4.3.4.2 Test equipment whose calibration has expired may be
used only with written request from the operations
manager and approval of the standards laboratory
manager. If this situation occurs, the test equipment
shall be calibrated as soon as feasible. If the
calibration data reflect out-of-tolerance values, the
maintenance performed with this test instrument shall
be evaluated (see Section 3.4.5). M&TE with an
expired calibration should be identified by a "rejected"
sticker and placed in a location to satisfy Section
3.4.3.2.1 of this procedure.
3.4.3.5 Control
3.4.3.5.1 The control process used for M&TE should be a well
established Recall Program containing, as a minimum
the following controls:
a) Total inventory listing
b) Calibration DUE notices - weekly
c) Individual automatic job opener and
information cards
d) Calibration OVERDUE notices - weekly
11
DOE-STD-1054-93
Section 7
e) Exception report for program errors
f) Scheduled and nonscheduled action history
reporting
3.4.3.5.2 Field use of other than "working" standards (i.e.,
primary/secondary) should be authorized only by the
accountable supervisor and then only under controlled
conditions and time limitations.
3.4.3.5.3 The application readiness of M&TE should be
regularly verified by applicable testing; and where
applicable, battery checks should be performed and
pass/fail criteria for each item should be clearly
indicated on each instrument. These readiness checks
shall not be a substitute for calibration activities.
3.4.3.5.4 Special-, limited-, or restricted-use M&TE shall be
clearly identified to describe its applicability.
3.4.3.6 Handling and Shipment of M&TE
During handling and shipment of M&TE devices M&TE integrity
shall be maintained. Packaging for shipment to the standards
laboratory should not conceal or destroy existing conditions of
equipment. Packaging after calibration and certification should be in
a manner which does not jeopardize equipment.
3.4.3.7 Traceability
M&TE shall be traceable from the field application, to the working
standard, secondary standard, primary standard, and NIST or other
nationally recognized standards. All calibration data should have both
forward and reverse traceability between the specific item and the
NIST.
12
DOE-STD-1054-93
3.4.4 Calibration of M&TE
3.4.4.1 M&TE should be calibrated using reference standards (secondary or
working) whose calibration has a known valid relationship to
nationally recognized standards or accepted values of natural physical
constants. If national standards do not exist, the basis for calibration
should be documented. The reference standard used should have an
accuracy at least four times greater than the device under test. If this
accuracy ratio cannot be met, analysis of the errors should be
estimated to provide a valid uncertainty of the calibration process.
3.4.4.2 M&TE Calibration Schedule and Interval
The standards laboratory, in conjunction with each department
manager responsible for M&TE, should establish and maintain a
calibration schedule for the M&TE (including primary, secondary, and
working standards. Specific items may be calibrated on a prior-to-use
basis rather than on a periodic basis. These devices shall be
controlled as specified in Sections 3.4.3.1 and 3.4.3.2 (tagging and
storage).
The schedule should normally be divided by application departments.
The M&TE assigned to a department should be listed in calibration
due date order by the unique identification number of Section 3.4.2.3
and noun name. Calibration dates for similar M&TE should be
staggered to ensure availability for day-to-day and peak needs. The
established scheduling organization should normally update and
distribute the schedule to the responsible department managers on a
quarterly basis. Department managers should ensure that M&TE
under their cognizance is calibrated in accordance with the master
schedule.
The calibration intervals for M&TE should be established by the
cognizant department managers. The initial interval should be based
on the inherent stability characteristics of the device, the rate/level and
purpose of expected use, manufacturer's recommendation, and
historical data for similar equipment. Calibration intervals may be
revised, based on a review of previous calibration results and
maintenance history. Revision of calibration intervals should require
the approval of the cognizant department manager. Revisions should
be documented on the Calibration Interval Change Authorization form
(Appendix C) which should be filed in the maintenance history.
Section 8
13
DOE-STD-1054-93
3.4.4.3 Calibration of M&TE - On Site
All M&TE on-site calibration should be done using plant-approved
procedures. These procedures should include the following items:
• identity of the equipment to be calibrated
• calibration equipment and reference standards to be used
• precautions and limitations
• checks, tests, measurements, acceptance tolerances, accuracy,
precision, range, and specific parameters
• step-by-step instructions regarding sequence, methods, data
sheet completion details, etc. (including as-found and as-left
conditions)
• special instructions, such as environmental conditions, when
appropriate
• reference standard used should have an accuracy at least four
times greater than the device under test. If this accuracy ratio
cannot be met, analysis of the errors should be estimated to
provide a valid uncertainty of the calibration process.
M&TE calibration data sheets should include provisions for recording
as-found data before any adjustments or repairs are made. Completed
data sheets should be reviewed, by the standards laboratory supervisor,
prior to release of the equipment to service to verify that acceptance
criteria are satisfied. When M&TE is found to be out of calibration
(gross error), an evaluation should be performed as described in
Section 3.4.5. Completed data sheets should be filed with the
maintenance history.
14
DOE-STD-1054-93
3.4.4.4 Calibration of M&TE - Off Site
M&TE may be calibrated by off-site organizations that have been
approved in accordance with on-site quality assurance department
procedures. Purchase orders or calibration requests to these
organizations should reference or include the following information:
• tolerance requirements
• calibration test data requirements
• any special environmental, handling, and shipping
requirements (see paragraph 3.4.3.6)
• calibration data to be supplied with the calibrated equipment
The off-site organization should also be required to provide immediate
verbal notification if the as-found data for an M&TE device does not
meet specified acceptance criteria. This notification is necessary to
ensure that the evaluation required by Section 3.4.5 is initiated as soon
as possible.
Upon return of the measuring and test equipment, the standards
laboratory supervisor shall ensure that the calibration data submitted
is reviewed, and an appropriate calibration sticker is attached.
3.4.5 M&TE Repairs/Out of Calibration
M&TE found to exceed required calibration tolerance or that has been subjected to
possible damage should be identified as rejected. When repairs are required to
standards or other measuring and test equipment they shall be re-calibrated to the
original requirements prior to being returned to normal service. When
calibration/certification is performed, the "as-found" condition and/or minor
adjustments to the measuring and test equipment shall be noted as part of the
equipment history information.
15
DOE-STD-1054-93
When M&TE is suspected or actually found to be inoperable, unreliable, defective,
or out of calibration, all data recorded since the previous calibration by affected
equipment shall be identified through the usage record described in Section 3.4.3.4.1
(see Appendix B). A prompt evaluation shall be performed to determine the need
for corrective action. This evaluation shall be documented on a Gross Error Report
(see Appendix D). The validity of all applications and data derived since the
previous calibration should be evaluated and dispositioned by the appropriate
owner/operator affected. The owner/operator of the affected equipment should
establish the nature and timing of corrective actions.
Section 9
3.4.6 Records
3.4.6.1 A maintenance history file should be kept for all M&TE. This file
should include the following, as appropriate:
• manufacturer's data, including model number and serial
number
• plant-unique identification number
• calibration interval and specifications
• history of calibrations, repairs, restrictions on use, and other
appropriate data
• calibration non-conformance evaluations (Section 3.4.5)
• usage record (Section 3.4.3.4.1)
• non scheduled actions
3.4.6.2 Records for lost, destroyed, unavailable, or removed-from-use M&TE
should be maintained in the history file.
3.4.6.3 Manufacturer's information manuals and supplemental bulletins should
be filed in accordance with document control procedures.
16
DOE-STD-1054-93
APPENDIX A
EXAMPLE OF "STICKERS" (TAGS) USED
IN M&TE CALIBRATION PROGRAM
A-1
DOE-STD-1054-93
APPENDIX A
Example of "Stickers" (Tags) Used in M&TE Calibration Program
CALIBRATION
Performed By: Date:
Expiration Date:
Figure 1
"Calibration Sticker"
I.D.No: Date:
RESTRICTED USE
By:
Figure 2
"Restricted Use Sticker"
REJECTED
I.D.No:
Date
Reason
By
Figure 3
"Rejected Sticker"
A-3
DOE-STD-1054-93
APPENDIX B
MEASURING AND TEST EQUIPMENT RECORD SHEET
B-1
DOE-STD-1054-93
APPENDIX B
MEASURING AND TEST EQUIPMENT RECORD SHEET
Plant
Facility Name
M&TE Description Serial No.
Model No. Date Last Calibration
Calibration Due Date
Identification of Component Calibrated/Tested
Date Device/loop Comp. Comp. Procedure No.
Calibrated/Tested Tag No. Ser. No. Job Request No.
B-3
DOE-STD-1054-93
APPENDIX C
CALIBRATION INTERVAL CHANGE AUTHORIZATION
C - 1
DOE-STD-1054-93
APPENDIX C
CALIBRATION INTERVAL CHANGE AUTHORIZATION
Instrument Number:
Manufacturer:
Model Number:
Change Calibration Interval from
to
Reason:
Requested by: Date:
Reviewed by: Date:
Approved by: Date:
C - 3
DOE-STD-1054-93
APPENDIX D
GROSS ERROR REPORT
D - 1
DOE-STD-1054-93
APPENDIX D
GROSS ERROR REPORT
FORM NO.
Plant
Facility Name
Instrument # SR#
Manufacturer Model #
Calibration Date Previous Calibration Date
Ranges Affected % Error Ranges Affected % Error
This instrument was involved in activities associated with the attached list of equipment
and/or jobs. Based on a review of the out-of-calibration condition and the affected list of
equipment, the checked action is required.
No Action Required:
instrument was not used during period (no list)
affected range(s) was not used
the error did not cause any test result/calibration to exceed equipment
allowable tolerances.
instrument would not function and failure time and cause were known,
providing reasonable assurance that the instrument was in calibration
prior to failure
other
D - 3
DOE-STD-1054-93
APPENDIX D (Cont.)
GROSS ERROR REPORT
Based on the Evaluation:
The equipment listed on the attached "retest" list requires
retest/rework. has been issued to ensure completion of
the required action
The equipment listed on the list was retested/reworked, and no further action
is required.
Other
Remarks
Review made by: Date:
Name
Approved by: Date:
Cognizant Supervisor
D - 4
DOE-STD-1054-93
APPENDIX E
CONTROL AND CALIBRATION OF MEASURING
AND TEST EQUIPMENT (M&TE)
SAMPLE LESSON PLAN
E - 1
DOE-STD-1054-93
APPENDIX E
CONTROL AND CALIBRATION OF MEASURING AND TEST EQUIPMENT (M&TE)
Section 10
SAMPLE LESSON PLAN
LESSON PLAN
1. The instructor should be familiar with the following background information:
a. Measuring and Test Equipment (M&TE) includes instruments, tools, and
devices used to calibrate or verify installed field equipment. Typically
M&TE equipment comes in two forms, primary and secondary standards. A
primary standard is traceable to the National Institute of Standards and
Technology (NIST) and is environmentally controlled at the owner facility.
The secondary standard is traceable to the primary and is used for field
calibration of installed equipment.
b. An effective M&TE program should contain the following elements:
• M&TE program based on traceable standards,
• controlled procedures used to calibrate M&TE,
• facilities established for control, storage, and issuance of M&TE,
• unique identification number for each piece of M&TE,
• any M&TE found out of calibration tolerance (gross error) receives a
timely evaluation to determine the condition of all the calibrations for
which it was used since its last successful calibration, and
• periodic reviews to determine that the control of M&TE is effective.
2. To teach this lesson, the following training housekeeping items are required:
a. Location for the training,
b. Approximately 30 minute time period for the training,
c. Notification of selected employees, and
d. A copy of the facility's M&TE program.
E - 3
DOE-STD-1054-93
APPENDIX E (Cont.)
3. This lesson has the following trainee enabling objective:
Explain the purpose and function of an M&TE program.
4. M&TE includes instruments, tools, and devices used to calibrate or verify installed
field equipment. Typically, M&TE equipment is defined as a primary or secondary
standard. A primary standard is traceable to NIST and is environmentally controlled
at the owner facility. A secondary standard is calibrated to the primary and used for
field calibration of installed equipment. Additional discussions on M&TE include
the following:
a. All calibrations of M&TE should be traceable to NIST. The primary standard
is traceable to NIST, the secondary to the primary, and field equipment to the
secondary. The use of primary standards for field use should be discouraged.
b. Calibration procedures should be used for all primary to secondary
calibrations. Only qualified technicians should perform these calibrations.
The interval of calibrations should be determined based on the manufacturer's
recommendations, the M&TE usage, and the M&TE's historical reliability.
c. Primary and secondary standards should be stored in an environmentally
controlled area. Issuance of M&TE should be controlled to prevent M&TE
damage due to wrong application, misuse of the M&TE by unqualified
workers, and field calibrations with out-of-tolerance M&TE.
d. A master list of all controlled M&TE should be maintained to include the
following:
• description of the equipment,
• discrete identification number,
• calibration procedures and periodic interval of calibration, and
• ranges of calibration and required accuracy.
5. Discuss with the trainees the facility's M&TE program.
E - 4
DOE-STD-1054-93
CONCLUDING MATERIAL
Review Activities: Preparing Activity:
DOE Field Offices DOE-NE-73
AD AL
DP CH Project Number:
EH ID
EM NV MNTY-0015
ER OR
NP RL
NS SR
RW SF
Area Offices
Amarillo
Brookhaven
Kansas City
Kirtland
Princeton
Facilities
ANL
BNL
LBL
PNL
PPPL
SNL
NV REECo.
NV EG&G
OR OSTI
WHC
EG&G
RF
SLAC
WSRC
U
ni
te
d
S
ta
Section 11
te
s
D
ep
ar
tm
en
t
of
E
ne
rg
y
O
ffi
ce
o
f
S
ci
en
tif
ic
a
nd
T
ec
hn
ic
al
I
nf
or
m
at
io
n
P
o
st
O
ffi
ce
B
ox
6
2
O
a
k
R
id
g
e
,
T
e
n
n
e
ss
e
e
3
7
8
3
1
FOREWORD
CONTENTS
1. INTRODUCTION
1.1 Purpose
1.2 Background
1.3 Application
2. DEFINITIONS
2. 1 Acronyms Used in this Standard.
2. 2 Calibration.
2. 3 Certification.
2. 4 Gross Error.
2. 5 Measuring & Test Equipment.
2. 6 Primary Standards.
2. 7 Recall Program.
2. 8 Secondary Standards.
2. 9 Standards Laboratory.
2.10 Systems Stations.
2.11 Working Standards.
3. CONTROL AND CALIBRATION OF MEASURING AND TEST EQUIPMENT (M&TE)
3.1 Discussion
3.2 Scope
3.3 Responsibilities
3.4 Guidelines
3.4.1 Procurement and Acceptance of M&TE
3.4.2 M&TE Identification
3.4.3 Control and Use of M&TE
3.4.4 Calibration of M&TE
3.4.5 M&TE Repairs/Out of Calibration
APPENDIX A EXAMPLE OF "STICKERS" (TAGS) USED IN M&TE CALIBRATION PROGRAM
APPENDIX B MEASURING AND TEST EQUIPMENT RECORD SHEET
APPENDIX C CALIBRATION INTERVAL CHANGE AUTHORIZATION
APPENDIX D GROSS ERROR REPORT
APPENDIX E CONTROL AND CALIBRATION OF MEASURING AND TEST EQUIPMENT (M&TE)
CONCLUDING MATERIAL