DOE-HDBK-1122-99 Module 2.19, Fundamental Academic Training Instructor's Guide Phase I; Module 2.19, Study Guide - Counting Room Equipment
Functional areas: Radiological Control, Technician Training, Study Guide, Counting Room Equipment
An overview of counters, scalers and associated equipment will describe the basic functions of counting equipment used to detect radiation activity. The RCT uses information from these counting instruments to identify and assess the hazards presented by contamination and airborne radioactivity and establish protective requirements for work performed in radiological areas. Stand-alone counters or scalers measure gross activity while spectroscopy systems perform spectrum analysis to identify and quantify activity from specific nuclides. The common uses of counting room equipment in various facilities will be discussed.
Unknown Block text
Document text
Text extracted from the attached file. Refer to the original document for the authoritative version.
Section 1
DOE-HDBK-1122-99
Module 2.19 Counting Room Equipment Study Guide
2.19-1
Course Title: Radiological Control Technician
Module Title: Counting Room Equipment
Module Number: 2.19
Objectives:
2.19.01 Describe the features and specifications for commonly used laboratory
counters or scalers:
a. Detector type
b. Detector shielding
c. Detector window
d. Types of radiation detected and measured
e. Operator-adjustable controls
f. Source check
g. Procedure for sample counting
2.19.02 Describe the features and specifications for low-background automatic
counting systems:
a. Detector type
b. Detector shielding
c. Detector window
d. Types of radiation detected and measured
e. Operator-adjustable controls
f. Source check
g. Procedure for sample counting
2.19.03 Describe the following features and specifications for commonly used
gamma spectroscopy systems.
a. Detector type
b. Detector shielding
c. Detector window
d. Types of radiation measured
e. Procedures
INTRODUCTION
An overview of counters, scalers and associated equipment will describe the basic
functions of counting equipment used to detect radiation activity. The RCT uses
information from these counting instruments to identify and assess the hazards presented
by contamination and airborne radioactivity and establish protective requirements for
work performed in radiological areas. Stand-alone counters or scalers measure gross
activity while spectroscopy systems perform spectrum analysis to identify and quantify
activity from specific nuclides. The common uses of counting room equipment in various
facilities will be discussed.
DOE-HDBK-1122-99
Module 2.19 Counting Room Equipment Study Guide
2.19-2
A variety of counting equipment is used. There are both manual and automated counting
systems. There is shielded equipment to measure radioactivity just above background
levels. There is equipment to measure gross counts of alpha, beta and gamma to
determine if surface contamination limits are met. There is equipment to measure the
energy spectrum for alpha and gamma radiation so that individual isotopes can be
identified and quantified (e.g. to determine if an alpha emitter is a plutonium isotope, a
uranium isotope or a radon daughter).
The counting systems use various types of detectors, including gas proportional counters
for alpha and beta radiation; sodium iodide, scintillation detectors for gamma
spectroscopy; zinc sulfide (ZnS) scintillation detectors for alpha radiation; liquid
scintillation for tritium and carbon 14; surface barrier (semiconductor) detectors for alpha
spectroscopy, lithium drifted germanium (GeLi semiconductor) detectors for gamma
spectroscopy, and high purity, germanium (HPGe semiconductor) detectors for gamma
spectroscopy.
The most common uses of the equipment are to count smears, swipes and air filters.
Nose swipes are also counted as one way to test if an individual has been exposed to
airborne radioactive contamination. Both workplace and stack emission air filters are
counted to measure the concentration of specific radionuclides (e.g. plutonium, and
uranium) and classes of radionuclides (e.g. mixed fission products).
References:
1. Radiation Detection and Measurement, Glenn F. Knoll
2. Basic Radiation Protection Technology, Daniel A. Gollnick
3. Operational Health Physics, Harold J. Moe
4. ANSI N323A
5. Various Manufacturer Technical Manuals
DOE-HDBK-1122-99
Module 2.19 Counting Room Equipment Study Guide
2.19-3
GENERAL PRINCIPLES
Section 2
A variety of counting room systems are used. The principles of these systems will be
discussed in general and then specific systems will be described.
Detector Type
When looking for low levels of radioactivity from alpha emitters (e.g. U, Pu, etc.) it is
important to minimize the background count rate from betas and gammas. The principle
used to accomplish this is pulse height discrimination. Betas have a range that is about
100 times greater than alphas, so alphas will deposit about 100 times as much energy in a
thin detector, producing a larger pulse than betas. Therefore alpha detectors are thin
(typically 1 mg/cm2) and use pulse height discrimination to distinguish alphas from betas.
Alpha detectors are generally either gas proportional counters, ZnS scintillators, or silicon
semiconductors.
Gamma spectroscopy requires good resolution to distinguish the different energy peaks.
GeLi or HPGe semiconductors give the best resolution, though NaI scintillators are also
used.
Detector Shielding
To reduce the background, shielding is often used. Betas can be shielded with aluminum
or plastic, while typical gamma shielding is a few inches of lead.
Detector Window
Since alphas have a short range the windows are thin, typically 1 mg/cm2 (or 0.25 mil
plastic). Some detectors have no window between the sample and the detector; in this
case there is a gas purge system for gas proportional counters, or a light tight housing for
scintillators. The alpha range is so short that self-shielding is often significant, e.g. an
alpha emitter buried in a filter may be shielded from the detector by the fibers.
Types of radiation
Some of the detectors discussed in objectives 1 and 2 are designed for alphas, some for
betas, and some will count both. Gamma spectroscopy is discussed in Objective 2.19.03.
Most nuclides emit more than one type of radiation, but beware of exceptions (like Be-7
or C-14).
Beta background is greater than alpha, so alpha detectors use pulse-height discrimination
to differentiate between alphas and betas.
DOE-HDBK-1122-99
Module 2.19 Counting Room Equipment Study Guide
2.19-4
2.19.01 Describe the features and specifications for commonly used laboratory
counters or scalers:
a. Detector type
b. Detector shielding
c. Detector window
d. Types of radiation detected and measured
e. Operator-adjustable controls
f. Source check
g. Procedure for sample counting
Some gammas will generally be detected in these detectors, but thin detectors have low
gamma efficiency, and lead shielding helps to reduce the gamma background still further.
Operator adjustable controls
Counting room systems have a timer to allow the operator to measure the number of
counts per minute (cpm). The most common count time is 1 minute, but the count time
can be selected by the operator.
Sources
National Institute of Standards and Technology (NIST) standard sources are used to
check the systems. Common sources are Pu-239 for alpha and Sr-90 for beta.
Procedures
Procedures generally include:
• background count
• source check
• sample count
• background subtraction
• divide by time to get cpm
• correct for 4 pi efficiency to get dpm
• record the data
LABORATORY COUNTERS OR SCALERS
In this section, specific laboratory counters or scaler systems are discussed, illustrating
the general principles discussed above.
DOE-HDBK-1122-99
Module 2.19 Counting Room Equipment Study Guide
2.19-5
Sample Holder SH-4 with HP210 and ESP
Section 3
The simplest system for counting smears is the portable contamination survey instrument,
the Eberline smart portable ESP with hand probe HP210.
Recall from lesson 2.17 that the HP-210 probe is a pancake GM detector with a thin (1.4
to 2 mg/cm2) window, suitable for detecting contamination from beta emitting nuclides,
and alphas above 3 MeV. The HP210T is shielded with tungsten to reduce gamma
background. The HP210AL is shielded with aluminum to reduce beta background.
The problem with using the HP210 for quantitative measurements (e.g. to satisfy release
criteria) is one of ensuring a precise geometry. The SH-4 sample holder solves this
problem by holding the sample in a fixed position directly under the HP210 detector.
Eberline Scintillation Alpha Counter SAC-4
The Eberline SAC-4 is a scintillation alpha counter. The scintillation phosphor is ZnS
powder on a plastic light pipe. The system is a self contained unit with the detector and
associated electronics housed in a single unshielded box. The detector and sample are
both in a light tight can, so no window is required between the ZnS detector and the
sample.
The system will accept samples up to 2 inches in diameter by 3/8 inches thick. (Self
shielding would be a major problem with samples this thick.) The sample holder in the
slide drawer is adjustable. It can be moved closer to the detector for thin samples.
However, the SAC-4 is calibrated with the sample holder in a certain position, so if the
sample holder is moved, the calibration is no longer valid.
The electronic package consists of the high voltage power supply used to power the
photomultiplier tube and determine its amplification, and a linear amplifier. The linear
amplifier output provides a 0 to 10 volt pulse signal to the discriminator that is set to
1.25V above the base line. Only pulses with amplitudes above 1.25V will be counted.
This will discriminate against betas because they will produce smaller pulses.
The output from the discriminator is counted by a six decade light emitting diode (LED)
readout. The timing circuit is synchronized to the line frequency (60 Hz) and provides
preset counting times from 0.1 to 50 minutes controlled by front panel switches. This
scaler can also be operated in a manual mode which will continue to count until reset by
the operator.
A Pu-239 source is used to check the system prior to each operating shift. Background
counts are conducted as a part of the performance check and to check for detector
contamination. The detector and sample drawer are easily removed for decontamination
if required.
DOE-HDBK-1122-99
Module 2.19 Counting Room Equipment Study Guide
2.19-6
The gross count rate is obtained by dividing total counts by the time in minutes.
Background counts (typically 0.3 cpm) are subtracted from gross counts to obtain net
counts per minute (cpm). The net count rate (cpm) is corrected for efficiency (as
described in lesson 2.03) to convert cpm to disintegrations per minute (dpm).
This counting system is used to obtain total activity and the procedures are followed as
described in the SAC-4 manual. Each background, source count, and sample count is
documented and kept on file.
NMC PC-5 and PC-55 Nuclear Measurement Corporation
Section 4
The PC-5 and PC-55 systems use gas flow proportional counters as the detectors. The
gas used in these detectors is P-10 (90% argon and 10% methane). The systems are self-
contained units with the detector and associated electronics housed in the same box. The
PC-55 is used to count both alpha and beta. The PC-5 may be manually adjusted to count
either alpha or beta. It is normally set to count alpha only. The determination of an alpha
count or a beta count is accomplished by pulse height discrimination.
No external shielding is used. Typical background for the unshielded detector is 2 cpm
for alphas, and 60 to 100 cpm for betas.
The PC-5 and PC-55 have identical detectors, 2.25 inch in diameter. They may be
installed with thin plastic windows with a thickness of 0.25 mil (0.00025 inch, 1 mg/cm2)
or they may be installed with no window. If there is no window, the operator must purge
with P-10 gas after inserting a sample and closing the gas tight door.
Front panel controls allow for pre-set gas purge times of 12, 36 and 144 seconds. If the
detector is installed with a thin plastic window, the normal procedure is to flow the gas
continually.
The sample to be counted is placed in a 2 inch diameter planchet and placed in the sample
drawer. The sample drawer then slides the sample under the detector. Should the
detector drawer or sample holder become contaminated during counting, it is a simple
task to remove the detector and drawer for decontamination.
The high voltage supply has a dual operating range of 300 - 1300 volts and 1300 - 2300
volts controlled from a front panel voltage potentiometer. The high voltage determines
the optimum setting to discriminate alphas from betas.
The count time is also set by front panel switches providing pre-set counting times in 0.1
minute increments up to 1000 minutes. In the automatic mode, the counter will count to
the pre-set time interval. In the manual mode, the counter will continue to count until
manually reset.
DOE-HDBK-1122-99
Module 2.19 Counting Room Equipment Study Guide
2.19-7
2.19.02 Describe the features and specifications for low-background automatic
counting systems:
a. Detector type
b. Detector shielding
c. Detector window
d. Types of radiation detected and measured
e. Operator-adjustable controls
f. Source check
g. Procedure for sample counting
Two sources are used to check the system for proper operation. The alpha source is Pu-
239 electroplated on a nickel disc. The beta source is 90Sr/Y90 (strontium 90 and its
daughter, Yttrium 90). These sources are traceable to NIST (National Institute of
Standards and Technology).
LOW-BACKGROUND AUTOMATIC SYSTEMS
In this section, several automatic counting systems are discussed. The principles are the
same as in Objective 2.19.01. The essential differences between the systems in
Objectives 2.19.01 and 2.19.02 are:
• complexity of electronics
• number of detectors or automated sample changing
• shielding to reduce background
Canberra 2400
The Canberra 2400 is a low background automatic counting system. The primary
detector is a gas flow proportional counter with a 2.25 inch diameter thin window, used to
count both alpha and beta activity. A second larger proportional counter, the guard
detector, is used to count background. The gas used is P-10 (90% argon and 10%
methane). The system may also incorporate a NaI scintillation detector, an option with
the Canberra 2400 systems, to simultaneously count gamma rays.
The sample detectors are surrounded by 4 inches of lead shielding to reduce background.
Typical background is 0.1 to 1 cpm alpha, 1 to 5 cpm beta, and 100 to 400 cpm gamma.
Section 5
Canberra 2400 systems are used primarily to count smears or filters. Gross counts for
each sample are processed in the computer and converted to dpm. Smear counts above
preset limits are highlighted and printed on a separate report.
DOE-HDBK-1122-99
Module 2.19 Counting Room Equipment Study Guide
2.19-8
Performance checks are performed daily or prior to system use. NIST traceable sources
of Pu-239 (plutonium), Sr-90 (strontium), and Tc-99 (technetium) are used. The system
should be calibrated regularly, or when chi-square values are outside the specified range
(see lesson 2.03 for a discussion of the chi-square test).
The system has an automatic sample changer with a dual stack that can handle up to 100
samples. One stack holds the samples to be counted and the other stack stores the
samples that have been counted.
Berthold LB770
The Berthold LB770 counting system is a low background semi-automatic counting
system. The system uses eleven P-10 gas flow proportional detectors; ten detectors are
used to count 10 radioactive samples simultaneously, the other detector is used to count
background radiation.
Each detector has a 2.25 inch diameter by 0.25 mil (0.00025 inch) mylar window. The
detector bay is shielded with 4 inches of epoxy coated lead. Typical backgrounds are 0.1
cpm alpha and 1 or 2 cpm beta. Typical counting efficiencies are 27% alpha and 42%
beta. The planchet is 0.25 inch deep, but a 0.25 inch thick sample would cause major
self-shielding problems (see lesson 2.03.01).
The Berthold systems are used primarily to count smears. Both alphas and betas are
counted simultaneously in each detector. Determination of alpha or beta activity is
accomplished by both pulse amplitude and pulse shape discrimination. The scalers in the
Berthold systems are similar to those used in the PC-55 but with more sophisticated
electronics that provide improved pulse shaping from the linear amplifier and better
discrimination of both pulse amplitude and pulse shape.
Pu 239 sources are used to check the system for alpha and 90Sr/90Y sources are used to
check for beta. These sources are traceable to NIST (National Institute for Standards and
Technology).
The Berthold system is controlled by a computer. Both alpha and beta counts received
from each sample are corrected for background and reported in one of three categories, to
alert the operator. If the count rate is below the minimum detectable activity (MDA, see
lesson 2.03) it falls into category 1. A count rate that falls within predetermined limits,
usually above MDA but below the limit for release to a controlled area (RadCon table
2.2) is category 2. And a count rate that is higher than the upper limit is category 3.
Background and efficiency data are collected for each detector, stored and used for
corrections. Pre-set count times are determined by the operator and put into the
computer. The count rate data from each detector is corrected and converted to dpm for
output to the printer.
DOE-HDBK-1122-99
Module 2.19 Counting Room Equipment Study Guide
2.19-9
2.19.03 Describe the following features and specifications for commonly used
gamma spectroscopy systems.
a. Detector type
b. Detector shielding
c. Detector window
d. Types of radiation measured
e. Procedures
Liquid Scintillation Counters, LSC e.g. Packard 2550
Tritium and C-14 emit such low energy betas that even a thin layer of air would stop the
betas. To detect this radiation, the sample must be in intimate contact with the detection
medium. This is achieved with a liquid scintillation system.
Section 6
A liquid scintillation counting system uses a "cocktail" that immerses the sample in the
counting medium to maximize the detection efficiency for low energy beta emitters. This
cocktail includes a liquid scintillator to convert the energy deposited by low energy betas
into light photons, which are then counted using photomultipliers.
The sample chamber, containing the sample vial and photomultiplier tubes, is light tight.
Since stray electrons can be spontaneously emitted from the photocathode, or by the
dynodes in the photomultiplier tube, two tubes are used with coincidence circuitry to
reduce this source of noise called "dark current". Typical background for beta is 20 cpm.
The LSC system is typically used to count tritium samples from swipes, water samples,
and oil samples (vacuum pumps). Tritium is also collected by drawing air samples
through silica-gel traps or glycol bubblers.
To calibrate the system, a series of cocktails with known amounts of tritium are prepared.
These sources are loaded into the first sample holder (a tray of 10 sample vials). The
computer program calculates the detector efficiency for each calibration source.
GAMMA SPECTROSCOPY
The instruments discussed in objectives 1 and 2 are designed to detect alphas and/or
betas, and make a gross count of total alpha and beta activity. In order to identify specific
radionuclides, the unique spectrum of energies particular to each radionuclide is used.
This technique is known as spectroscopy.
DOE-HDBK-1122-99
Module 2.19 Counting Room Equipment Study Guide
2.19-10
Alpha emitters (e.g. Th, U, Pu, Am and their daughters) have characteristic alpha
energies, but alpha spectroscopy, detecting the alphas directly, is not optimal, because the
energy loss of alpha particles between the sample and the detector smears the energy
spectrum (see lesson 2.18.03). Gamma spectroscopy looks for the characteristic spectrum
of gammas from the radioactive decay.
Gamma spectroscopy usually uses germanium detectors (GeLi or HPGe) because the
good resolution obtained with these detectors enables gammas with nearly the same
energy to be distinguished or resolved.
EG&G Ortec Gamma X
The Gamma X Spectroscopy system uses an HPGe coaxial photon detector to perform
gamma and x-ray spectroscopy in the energy range from 3 keV to 10 MeV.
Detector Type
The detector is made of n-type high purity germanium semiconductor (HPGe). A 30 liter
dewar of liquid nitrogen (LN2) is used to cool the detector.
Detector Shielding
The detector is shielded by 4 inches of pre World War II steel. This steel is used when a
low background is desired as it was manufactured before radioactive fallout (artificial
radioactivity, lesson 1.06.03) from nuclear weapons appeared in trace quantities. A
sample holder inside the shield allows the sample to be positioned at distances from less
than 1 cm up to 40 cm from the detector end cap.
Detector Window
The detector window is 0.5 mm thick beryllium.
Types of Radiation Measured
The gamma spectrometer is designed to detect gammas and x-rays from alpha emitting
nuclides, and sort the data in a multi channel analyzer to produce a spectrum that is
characteristic of the nuclide. The peaks in the spectrum are close together, so excellent
resolution is required to distinguish the peaks. Typical resolution from a germanium
semiconductor detector (HPGe or GeLi) is better than 1%, which means that if the photon
energy is 100 keV, the width of the peak is less than 1 keV. Photons from two different
nuclides that are 1 keV apart will be seen as two distinct peaks.
Section 7
DOE-HDBK-1122-99
Module 2.19 Counting Room Equipment Study Guide
2.19-11
Procedures
Energy and efficiency calibrations are obtained using two different sources that are NIST
standards. These are mixed sources that contain several gamma emitting nuclides. One
source contains isotopes of americium (Am), antimony (Sb), and Europium (Eu). The
second mixed source contains isotopes of cadmium (Cd), cerium (Ce), cobalt (Co),
strontium (Sr), tin (Sn), cesium (Cs), and yttrium (Y). The energy and efficiency
calibration values are then used by the analysis software.
Specific procedures are written to direct the operator through the sample and computer
setup, and the computer analysis. The original copy of the results is kept on file for 1
year and then archived for 75 years.
SUMMARY
This lesson has discussed the detector, shielding, window, types of radiation detected, and
procedures for counting room equipment. This knowledge is important to ensure accurate
and consistent counting room data for the assignment of proper radiological controls.
SUMMARY
HP210, SH-4 and ESP Laboratory Counter
Detector type: GM
Detector shield: Tungsten for gamma, or aluminum for beta
Detector window: mica 1.4 to 2 mg/cm2
Radiation detected: beta
Controls: ESP (see lesson 2.17)
Source check: Sr-90
Eberline SAC–4 Laboratory Counter
Detector type: ZnS(Ag) scintillator
Detector shield: None
Detector window: None
Radiation detected: alpha
Controls: Timer: 0.1 to 50 minutes
Source check: NIST traceable Pu-239 source
DOE-HDBK-1122-99
Module 2.19 Counting Room Equipment Study Guide
2.19-12
NMC PC–5, PC-55 Laboratory Counter
Detector type: Gas flow proportional counter (P-10 gas)
Detector shield: None
Detector window: None, or optional 0.25 mils plastic (1 mg/cm2)
Radiation detected: alpha
Controls: Timer 0.1 minute to 1000 minutes
Source check: NIST traceable Pu-239 source
Canberra 2400 Low-Background Automatic
Detector type: Gas flow proportional detector (P-10 gas).
Larger guard detector for background.
Optional NaI detector for gammas.
Detector shield: 4 inches lead
Detector window: 0.25 mil mylar (1 mg/cm2)
Radiation detected: alphas and betas
Optional NaI detector for gammas
Source checks: Pu-239, Sr-90, and Tc-99.
Berthold LB770 Low-Background Automatic
Detector type: 10 gas flow proportional counters (P-10 gas).
Plus one detector to count background.
Detector shield: 4 inches of epoxy coated lead
Detector window: 0.25 mil mylar (1 mg/cm2)
Radiation detected: alphas and betas
Source check: NIST traceable Pu-239 calibration source for alpha and 90Sr/Y90 for
beta
Liquid Scintillation Counter
Packard 2550 Low-Background Automatic
Detector type: Liquid scintillation
Detector shield: none
Detector window: none (Light tight housing)
Radiation detected: Low energy beta from Tritium (or C-14)
Source checks: LANL calibration sources, with water as quench agent.
EG&G Ortec Gamma-X Gamma Spectroscopy
Detector type: HPGe (high purity germanium)
Detector shield: 4 inches pre WW-II steel or 4 inches of coated lead.
Detector window: Beryllium window 0.5 mm thick
Radiation detected: gamma and x-ray photons from 3 keV to 4MeV
Source checks: NIST traceable mixed sources:
DOE-HDBK-1122-99
Module 2.19 Counting Room Equipment Study Guide
2.19-13
GLOSSARY:
cocktail: mixture of liquid scintillation chemicals and sample
discriminator: electronic device that discriminates against small pulses, e.g. to distinguish
alphas from betas.
gamma spectroscopy: the use of gamma spectra to identify radionuclides by their
characteristic gamma emissions.
multi channel analyzer (MCA): combination of many SCAs, each connected to a scaler
channel, to produce a spectrum
resolution: measure of the ability of a system to separate nearby peaks in a spectrum;
measure of the widths of the peaks.
single channel analyzer (SCA): combination of a lower level discriminator and an upper
level discriminator to select only pulses between the two levels (e.g. to select betas but
reject small pulse height noise and large pulse height alphas).
DOE-HDBK-1122-99
Module 2.19 Counting Room Equipment Study Guide
2.19-14
This page intentionally left blank.
Module Number: 2.19