DOE-HDBK-1122-99 Module 2.16, Fundamental Academic Training Instructor's Guide Phase I; Module 2.16, Study Guide - Radiation Survey Instrumentation
Functional areas: Radiological Control, Technician Training, Study Guide, Radiation Survey Instrumentation
External exposure controls used to minimize the dose equivalent to personnel are based on the data taken with portable radiation survey instruments. An understanding of these instruments is important to ensure the data obtained are accurate and appropriate for the source of radiation. This lesson contains information about widely used portable radiation survey instruments.
Unknown Block text
Document text
Text extracted from the attached file. Refer to the original document for the authoritative version.
Section 1
DOE-HDBK-1122-99
Module 2.16 Radiation Survey Instrumentation Study Guide
2.16-1
Course Title: Radiological Control Technician
Module Title: Radiation Survey Instrumentation
Module Number: 2.16
Objectives:
2.16.01 List the factors which affect an RCT's selection of a portable radiation
survey instrument, and identify appropriate instruments for external
radiation surveys.
i 2.16.02 Identify the following features and specifications for ion chamber
instruments used at your facility:
a. Detector type
b. Instrument operating range
c. Detector shielding
d. Detector window
e. Types of radiation detected/measured
f. Operator-adjustable controls
g. Markings for detector effective center
h. Specific limitations/characteristics
i 2.16.03 Identify the following features and specifications for high range
instruments used at your facility:
a. Detector type
b. Instrument operating range
c. Detector shielding
d. Detector window
e. Types of radiation detected/measured
f. Operator-adjustable controls
g. Markings for detector effective center
h. Specific limitations/characteristics
i 2.16.04 Identify the following features and specifications for neutron detection and
measurement instruments used at your facility:
a. Detector type
b. Instrument operating range
c. Types of radiation detected/measured
d. Energy response
e. Operator-adjustable controls
f. Specific limitations/characteristics
DOE-HDBK-1122-99
Module 2.16 Radiation Survey Instrumentation Study Guide
2.16-2
INTRODUCTION
External exposure controls used to minimize the dose equivalent to personnel are based
on the data taken with portable radiation survey instruments. An understanding of these
instruments is important to ensure the data obtained are accurate and appropriate for the
source of radiation. This lesson contains information about widely used portable
radiation survey instruments.
Many factors can affect how well the measurement reflects the actual conditions, such as:
• Selection of the appropriate instrument based on type and energy of radiation,
radiation intensity, and other factors.
• Correct operation of the instrument based on the instrument operating
characteristics and limitations.
• Calibration of the instrument to a known radiation field similar in type, energy
and intensity to the radiation field to be measured.
• Other radiological and non-radiological factors that affect the instrument
response, such as RF fields, radioactive gases, mixed radiation fields, humidity
and temperature.
References:
1. Radiation Detection and Measurement, Glenn F. Knoll
2. Basic Radiation Protection Technology, Daniel A. Gollnick
3. Operational Health Physics, Harold J. Moe
4. ANSI N323A
DOE-HDBK-1122-99
Module 2.16 Radiation Survey Instrumentation Study Guide
2.16-3
NOTE: The text is provided for some commonly used instruments. The facility
must adjust text as necessary for instruments used at each site. Text added
for specific instruments used at the facility must, at a minimum, cover
material required by the objectives.
2.16.01 List the factors which affect an RCT's selection of a portable radiation
survey instrument, and identify appropriate instruments for external
radiation surveys.
FACTORS AFFECTING INSTRUMENT SELECTION
As discussed, the selection of the proper instrument is critical to ensure the data obtained
are accurate and appropriate. The instrument is selected based on the characteristics and
specifications for that instrument as compared to the required measurements. Several
factors should be considered when selecting the instrument.
Section 2
• Type of Data Required
Distinguish clearly between external radiation surveys (lesson 2.16) and
contamination monitoring (lesson 2.17). External radiation surveys require an
instrument that reads R/hr, mR/hr, rem/hr, mrem/hr, etc., rather than counts per
minute, etc.
• Measurement of the True Dose Equivalent
Ion chambers (which read current instead of counting pulses) have the flattest
energy response. Ion chambers are closest to being tissue equivalent. Generally,
the best choice for external beta-gamma surveys is an ion chamber.
• Type of Radiation to be Measured
Ion chambers measure beta and gamma. For neutrons, choose a rem ball (NRD).
Alphas are not measured in an external radiation survey, since they do not
penetrate the skin (7 mg/cm2, see lesson 1.07.10).
• Intensity of the Radiation (exposure or dose rate)
For high radiation fields (>5 R/hr) use an extendible instrument (Teletector) if this
is "reasonably achievable" (ALARA).
DOE-HDBK-1122-99
Module 2.16 Radiation Survey Instrumentation Study Guide
2.16-4
• Energy of the Radiation to be Measured
Low energy radiation will not penetrate either the skin or the window of most
external radiation instruments. GM detectors over-respond to low energy
gammas. Most instruments under-respond to high energy neutrons.
• Environmental Factors
Ion chambers are usually vented to air, so radioactive gases or high humidity
affect the instrument response.
• Procedures
If all else fails, read the instructions!
Preoperational Check
Once the proper type of instrument has been identified, a pre-operational check is
essential and must be performed in accordance with appropriate procedures.
• Physical Damage
Perform a physical inspection of the instrument by checking for obvious physical
defects or damage, especially of the probe, and replace the probe or cable if
necessary.
• Calibration Sticker
Verify the instrument is calibrated and has not exceeded the calibration due date.
• Battery
Perform a battery check to verify the battery condition is within the acceptable
range. Change the batteries if necessary.
• Zero
Perform a zero adjustment for the meter needle, if applicable (e.g. for ion
chambers).
• Source Check
Perform a source response check as required by the procedures.
DOE-HDBK-1122-99
Module 2.16 Radiation Survey Instrumentation Study Guide
2.16-5
Instrument Selection General Principles
To ensure the proper selection and operation of instruments, the instrument operator must
understand the operating characteristics and limitations of each instrument available for
use.
There are general principles which apply to the specific instruments described in the
following sections.
• Detector Type
Ion chambers have the flattest energy response. Ion chambers are closest to being
tissue equivalent.
GM detectors over-respond to low energy gammas
Special detectors are used for neutrons
• Instrument Operating Range
External radiation measuring instruments read in R/hr, rad/hr, or rem/hr. In
contrast, instruments designed for measuring contamination read in cpm.
Extendible instruments are generally appropriate for high radiation fields.
• Detector Shielding
Large amounts of shielding are not practical with a portable instrument, but some
probes incorporate a small amount of shielding to reduce the background. Many
external radiation survey instruments incorporate a sliding "beta shield". Note
that this also shields low energy gammas.
• Detector Window
Section 3
External radiation instruments generally have windows that are about as thick as
human skin (7 mg/cm2). The reason for this is: if the radiation does not penetrate
this window then it does not penetrate skin, and so it does not contribute any
external dose. In contrast, contamination monitoring instruments have thinner
windows.
• Types of Radiation Detected/Measured
Ion chambers have a flat energy response for gammas. Ion chambers are closest
to being tissue equivalent. They are also good for betas, but a correction factor
may be needed.
DOE-HDBK-1122-99
Module 2.16 Radiation Survey Instrumentation Study Guide
2.16-6
Tube shaped GM detectors are designed so that the walls are close to the detector
gas. Gamma interactions in the walls are important. A well designed detector wall
can partially compensate for the over-response to low energy gammas. They are
designed primarily for gammas, and also measure betas if the window is not too
thick.
Pancake shaped GM detectors have side walls separated from the gas. They are
good for betas, but have a low efficiency for gammas because very few gammas
hit the side walls.
Gas proportional detectors distinguish between alphas and betas. They often
discriminate against (reject) betas and gammas.
ZnS scintillation detectors only detect alphas.
NaI scintillation detectors are generally used for gammas.
Neutron detectors are very specialized.
• Operator Adjustable Controls
Portable instruments generally have a battery check.
Ion chambers generally have a zero adjustment.
• Markings for Effective Detector Center
External radiation surveys are generally taken at 30 cm (except for transportation,
see lesson 2.12). It is not always obvious what point on the detector should be 30
cm from the source, so most detectors mark the effective center.
The effective center of the detector, as defined in ANSI N323, is the point within
the detector that produces, for a given set of irradiation conditions, an instrument
response equivalent to that which would be produced if the entire detector were
located at that point. The effective center can be thought of as the point in the
detector where the measurement of the radiation intensity is taken. Portable
radiation survey instruments are calibrated in a uniform field of radiation larger
than the volume of the detector, so that the same radiation intensity is seen
throughout the detector. Therefore, the reading "taken" at the effective center
represents the rate value in all portions of the detector. If the radiation field over
the whole detector is not uniform, the exposure rate will not be uniform over the
entire detector volume. For non-uniformly irradiated detectors, the displayed
value, as "taken" at the effective center, will not reflect the actual exposure rate
value and a correction factor may be appropriate.
DOE-HDBK-1122-99
Module 2.16 Radiation Survey Instrumentation Study Guide
2.16-7
2.16.02 Identify the following features and specifications for ion chamber
instruments used at your facility:
a. Detector type
b. Instrument operating range
c. Detector shielding
d. Detector window
e. Types of radiation detected/measured
f. Operator-adjustable controls
g. Markings for detector effective center
h. Specific limitations/characteristics
EBERLINE RO-2 and RO-3
The Eberline RO-2, RO-3 series of instruments are portable, air-vented ion chamber
instruments used to detect and measure gamma, X-ray, and beta radiation. Technical
specifications for the RO3 are similar to the RO2.
Section 4
Detector Type
The ion chamber is a phenolic, or plastic cylinder of 3 in. diameter and 12.7in3 (208cm3)
volume, with one end covered by a Mylar window. The fill gas is air, vented to
atmosphere through a desiccant pack.
The ion chamber detector is closer to tissue-equivalent than most types, allowing the
instrument to assess the exposure rate to human tissue. The detector is approximately
tissue equivalent because the materials used for construction have an effective atomic
number Z close to that of tissue at 7.5. "Tissue equivalent" means that the detector
responds the same as human soft tissue. No detector is perfectly tissue equivalent, but a
well designed ion chamber is close enough for most work.
Although the detector is not as sensitive as a GM, it is the detector of choice for assessing
exposure because of its close correlation to the energy deposited in human tissue by
radiation.
The RO-2 series instruments are operated in the current mode, which is the mode that
averages the individual pulse heights per unit time. Individual pulse information is lost;
therefore, the electrical signal will not supply information about the type and energy of
the individual radiation interactions. However, small pulses, which would be lost in the
pulse mode, are averaged along with the other interactions.
DOE-HDBK-1122-99
Module 2.16 Radiation Survey Instrumentation Study Guide
2.16-8
Instrument Operating Ranges
The instrument range of the Model RO-2 is 0 - 5000 mR/hr. The readings are expressed
in roentgen, since the measurement is made in air.
The settings are as follows:
RO-2 Ranges
0-5 mR/hr
0-50 mR/hr
0-500 mR/hr
0-5,000 mR/hr
Detector Shielding
The sliding beta shield is made of phenolic as follows:
• RO-2 shield: 400mg/cm2 (1/8 in. thick) mounted on the case.
The active volume of the detector is shielded from the side by the detector wall and the
instrument case, and from the bottom by the movable beta shield and two layers of
window. The detector wall is 200mg/cm2 and the 0.13 cm aluminum case is about 345-
mg/cm2.
Detector Window
The materials and density-thickness value of the two windows, one on the case and one
on the detector, for the Model RO-2 and RO3 are as follows:
• RO-2 windows: 7mg/cm2 total; two Mylar windows 3.5mg/cm2 (1 mil) each.
• RO-3 window: 3.5mg/cm2 total; one window of 1 mil Mylar.
Types of Radiation Detected/Measured
The RO-2 instrument is designed to measure gamma, X-ray, and beta radiation but will
respond to (not measure) neutron radiation. Although an ionization chamber would
respond to alpha radiation, the Mylar windows and the air gap between the two windows
eliminates any possibility of an alpha response.
DOE-HDBK-1122-99
Module 2.16 Radiation Survey Instrumentation Study Guide
2.16-9
The RO-2 measures photon radiation within ±20% for photon energies from 12keV to 7-
MeV (beta shield open). The minimum energy increases to 25keV if the shield is closed,
and to about 40keV through the side of the instrument. Because of the thinner window,
the RO-3 measures photons from 8-keV.
The RO-2 measures beta radiation >70keV with the beta shield open. A beta correction
factor may be appropriate in some situations.
Operator-Adjustable Controls
RO-2 range switch with OFF, ZERO, and BATT checking positions.
ZERO position works in conjunction with ZERO knob to electronically zero the meter.
BAT1 and BAT2 positions check the two batteries used to power the instrument circuitry.
Section 5
Markings for Detector Effective Center
The Effective Center markings on the RO-2 are the "dimples" or depressions on the sides
and front of the instrument case.
Specific Limitations\Characteristics
The response time for the RO-2 series of instruments is 5 seconds to reach 90% of the full
value.
High humidity or moisture can cause leakage currents in the detector and cause erratic
meter readings. The detector is vented through a silica gel desiccant, or drying medium,
contained in a plastic box. The desiccant can become saturated and will need replacement
if the crystals start to turn clear or pink instead of the normal blue color.
The detector is vented to atmosphere; therefore, any change in atmospheric density
changes the air density in the detector. An increase in atmospheric pressure will cause an
increase in air density in the detector and cause a higher response. If the RO2 is calibrated
in Los Alamos (7000 ft) and then used at sea level, the response will be 30% high. A
change in response of about 10% will occur if the instrument was calibrated at room
temperature and used in an environment that is different by about 50 bF.
Because the detector is vented to atmosphere, radioactive gases could enter the detector
and cause a reading.
BICRON RSO-50 AND RSO-500 INSTRUMENTS
The Bicron RSO-50 and RSO-500 instruments are portable air-vented ion chamber
instruments used to detect and measure gamma, X-ray and beta radiation. The Bicron
DOE-HDBK-1122-99
Module 2.16 Radiation Survey Instrumentation Study Guide
2.16-10
RSO series of instruments are very similar in design and construction to the Eberline RO-
2 series of instruments.
Detector Type (identical for both models)
• Operated as an ionization chamber.
• A phenolic, or plastic, cylinder of 3 in. diameter and 12.7 in3 (208 cm3) volume
with one end open but covered by a Mylar window.
• Fill gas - air (vented to atmosphere through a silica gel desiccant pack).
The Bicron RSO series instruments are operated in the current mode, or the mode that
averages the individual pulse heights per unit time.
Instrument Operating Range
The instrument ranges of the two models are as follows:
RSO-500 Ranges RSO-50 Ranges
0-0.5 R/hr 0-50 mR/hr
0-5 R/hr 0-500 mR/hr
0-50 R/hr 0-5 R/hr
0-500 R/hr 0-50 R/hr
Detector Shielding
The active volume of the detector is shielded from the side by the detector wall and the
instrument case and from the bottom by the movable beta shield and two layers of
window. Detector wall is 200 mg/cm2 and the 0.13 cm aluminum case is about
345 mg/cm.
Detector Window
The materials and density-thickness value of the two windows, one on the case and one
on the detector, are the same for both models.
• RSO windows - 7 mg/cm2 total, both windows are Mylar of 5 mg/cm2 each.
The sliding beta shield is made of phenolic and the density-thickness value is the same for
both models.
DOE-HDBK-1122-99
Module 2.16 Radiation Survey Instrumentation Study Guide
2.16-11
• RSO shield - 400 mg/cm2 (1/8 in. thick) and is mounted externally on the case.
Types of Radiation Detected/Measured
The Bicron RSO series of instruments are designed to measure gamma, X-ray and beta
radiation but will detect (not measure) fast neutron radiation. The instruments will read
approximately 10%, in mR/hr, of the true neutron field, in mrem/hr. Like the Eberline
RO-2, the Bicron RSO series instruments will not respond to alpha radiation because the
alpha particles are shielded before they reach the detector. The energy response of the
two models is identical. Both models measure photon radiation within +20% for photon
energies from 12 keV to 7 MeV (beta shield open). The minimum energy increases to 25
keV if the shield is closed, and to about 40 keV through the side of the instrument. Both
models measure beta radiation >70 keV.
Section 6
Operator-Adjustable Controls
RSO-500 range switch with OFF, ZERO, and BATT positions.
• Switch ranges labeled as 0.5, 5, 50, and 500 R/hr.
• ZERO position works in conjunction with ZERO knob to electronically zero the
meter.
• BAT position checks the two batteries used to power the instrument circuitry and
detector bias.
• OFF position turns the instrument off.
RSO-50 range switch is the same but is labeled 50 and 500 mR/hr and 5 and 50 R/hr.
Markings for Detector Effective Center
The effective center markings on both Bicron models are the stamped circles with a plus
sign in the circle and are located on the sides and front of the instrument case. If the
radiation field over the whole detector is not uniform ( such as from surface
contamination, radiation streaming, or from a small point source) the displayed value may
need to be corrected.
Specific Limitations/Characteristics
The response time varies between the two models of Bicron instruments available.
RSO-500 - approximately 10 sec from 0-90% of the final reading.
RSO-50 - approximately 5 sec from 0-90% of the final reading.
DOE-HDBK-1122-99
Module 2.16 Radiation Survey Instrumentation Study Guide
2.16-12
Correction factors may be needed when the radiation field is not uniform over the entire
detector. High humidity or moisture can cause leakage currents in the detector and cause
erratic meter readings.
• The detector is vented through a desiccant, or drying medium, contained in a
plastic box.
• The desiccant can become saturated and will need replacement if the crystals start
to turn clear or pink.
Like the Eberline RO-2, the detector is vented to atmosphere; therefore, any change in
atmospheric density changes the air density in the detector.
• An increase in temperature will lower the air density in the detector and cause a
lower response.
• An increase in atmospheric pressure will cause an increase in air density in the
detector and cause a higher response.
• Tables are provided in the technical manuals for correcting the instrument
response due to changes in pressure or temperature.
• A change in response of about 10% will occur if the instrument was calibrated at
room temperature and used in an environment that is different by about 50 bF.
Because the detector is vented to atmosphere, radioactive gases can enter the detector and
cause a reading.
VICTOREEN MODEL 450B
The Victoreen 450B is a portable, general purpose, ion chamber survey instrument which
uses microprocessor and LCD (liquid crystal display) technology.
Detector Type
Operated as an ionization chamber. A Bakelite, or plastic, cylinder of 200 cm3 volume
with one end open but covered by a Mylar window. The fill gas is air (vented to
atmosphere through a desiccant pack). The ion chamber detector is designed as tissue-
equivalent, allowing the instrument to accurately assess the exposure rate to human
tissue. The Victoreen 450B is operated in the current mode as are most ion chambers.
DOE-HDBK-1122-99
Module 2.16 Radiation Survey Instrumentation Study Guide
2.16-13
Instrument Operating Ranges
• Overall range is 0-50 R/hr.
The instrument is autoranging, or automatically changes scales as required for the
instrument reading, and has the following scales:
450B Scales
0-5 mR/hr
0-50 mR/hr
0-500 mR/hr
0-5 R/hr
0-50 R/hr
Detector Shielding
Section 7
The active volume of the detector is shielded from the side by the detector wall and the
instrument case and from the bottom by the movable beta shield and windows. The
detector wall is 200 mg/cm. The sliding beta shield is made of Bakelite, which is a type
of plastic and the density-thickness value is 440 mg/cm.
Detector Window
The two detector windows, one on the detector and one on the case, are made of 1.7 mil
Mylar for a total of 3.4 mg/cm2.
Type of Radiation Detected/Measured
The 450B instrument is designed to measure gamma, X-ray, beta and alpha radiation but
will detect (not measure) fast neutron radiation. The instruments will read approximately
10%, in mR/hr, of the true neutron field, in mrem/hr.
Energy response. Photon energy response (+20%) is about 20 keV for slide open, 40 keV
for slide closed, and 50 keV from the side. Beta energies >32 keV can be measure. The
alpha response is limited to energies >4 MeV and only if the detector to source distance is
less than the alpha range in air.
DOE-HDBK-1122-99
Module 2.16 Radiation Survey Instrumentation Study Guide
2.16-14
Operator-Adjustable Controls
Only three external controls are available on the 450B: the ON/OFF switch, the MODE
switch and the meter light button. The Mode switch is used during calibration and is not
enabled for operator use. The ON/OFF switch turns the instrument on and off. The
instrument is autoranging and will change the bar graph, digital value and scale markings
as appropriate for the exposure rate value. The instrument has an "autozero" feature that
eliminates any need for an external zero control. If the batteries are low, then the
instrument will display a LOW BAT message. A button switch is provided in the handle
for turning the meter face light on and off.
Markings for Detector Effective Center
The effective center markings on the 450B are the painted-white depressions in the plastic
case and are located in the front and on the sides.
Specific Limitations/Characteristics
The response times to 90% of the final value for the 450B instrument are as follows,
assuming that a step increase or decrease in the rate does not cause a range change:
450B Response Times
0-5 mR/hr 8 sec
0-50 mR/hr 5 sec
0-500 mR/hr 2 sec
0-5 R/hr 2 sec
0-50 R/hr 2 sec
Geotropism, or the effect of gravity on the instrument, causes no greater than a +1% of
full scale change from the actual value. Correction factors must be applied when the
radiation field is not uniform over the entire detector, such as for surface contamination
beta dose rates. High humidity or moisture could cause leakage currents in the detector
and cause erratic meter readings. The detector is vented through a desiccant, or drying
medium, contained in a plastic cylinder. The desiccant could become saturated and will
need replacement if the crystals start to turn clear or pink. The atmospheric vent on the
case has a rubber bladder to allow for changes in temperature and pressure but prevents
the free flow of air into and out of the detector casing. The rubber bladder minimizes the
effects of high humidity environments and radioactive gases. The detector is vented to
atmosphere; therefore, any change in atmospheric density changes the air density in the
detector. An increase in temperature will lower the air density in the detector and cause a
lower response. An increase in atmospheric pressure will cause an increase in air density
DOE-HDBK-1122-99
Section 8
Module 2.16 Radiation Survey Instrumentation Study Guide
2.16-15
2.16.03 Identify the following features and specifications for high range
instruments used at your facility:
a. Detector type
b. Instrument operating range
c. Detector shielding
d. Detector window
e. Types of radiation detected/measured
f. Operator-adjustable controls
g. Markings for detector effective center
h. Specific limitations/characteristics
in the detector and cause a higher response. The value of the changes due to temperature
and pressure are similar to those of other air-vented ion chambers.
EBERLINE TELETECTOR
The Eberline Teletector is an extendible, telescoping-rod instrument designed with two
Geiger-Mueller (GM) detectors for the measurement of photon exposure rates and
detection of beta radiation.
Detector Types
Both detectors are sealed GM tubes with halogen-quenched argon fill gas contained in an
energy compensating case. Energy compensation is required in GM detectors to reduce
the over response to low energy photons.
The low range detector is the largest of the two detectors and is located at the end of the
detector housing. The low range detector is used for the three lowest ranges on the
instrument.
The high range detector is the small cylinder in the detector housing, and is used for the
upper two scales.
The GM detectors are very sensitive; however, they lack the direct correlation to energy
deposited and are not as accurate as ion chamber instruments for assessing exposure rates.
The Teletector instrument is operated in the pulse mode, or the mode that counts each
individual pulse. Since any ionization in a GM tube causes the same large pulse, any
radiation interaction in the detector will be counted. All the pulses are of the same large
size regardless of the energy or type of radiation; therefore, all information on the type
and energy of the radiation is lost.
DOE-HDBK-1122-99
Module 2.16 Radiation Survey Instrumentation Study Guide
2.16-16
Instrument Operating Ranges
The instrument range is 0 - 1000 R/hr.
The analog Teletector has five settings. The three lower settings utilize the large GM
detector and the two upper settings utilize the smaller GM detector.
Teletector Settings
0-2 mR/hr
0-50 mR/hr
0-2 R/hr
0-50 R/hr
0-1000 R/hr
Detector Shielding
The two detectors are shielded by layers of lead and fiber to compensate for the GM over-
response to low-energy photons. The high-range detector is partially shielded by the low-
range detector.
Detector Window
The low-range detector has a 30mg/cm2 mica window and a rubber cap to protect the
window.
Types of Radiation Detected/Measured
The Eberline Teletector will measure gamma and X-ray radiation and responds to (but
does not measure) beta radiation. According to the manufacturer, beta response is not
accurate. Alpha response is eliminated by the thicker window and casing. Neutron
response is insignificant due to the lower probability of interaction in the small detectors.
The Teletector measures photon radiation >80 keV. Lower energy photons are
attenuated in the detector window. Beta particles >160 keV can be detected but not
measured.
Operator-Adjustable Controls
The only control is the range switch with OFF and B (battery check) positions.
DOE-HDBK-1122-99
Module 2.16 Radiation Survey Instrumentation Study Guide
2.16-17
Detector Effective Center Markings
The effective center of both detectors is indicated by the machined grooves in the detector
housing, with the groove closest to the beta window indicating the low-range detector.
Section 9
Specific Characteristics and Limitations
Response time for the instrument is approximately 1 second to 90% of full scale.
The sealed detectors do not require correction factors for temperature or pressure. The
sealed detectors do not experience problems with humidity or radioactive gases entering
the detector.
Audible indication is available only through the speaker jack; no internal speaker is
installed.
For GM detectors, the possibility exists that the detectors become saturated in very high
radiation fields. Some GM detector instruments will read zero if the detector becomes
saturated. The manufacturer states (on the title page of the 6112B brochure) that the
Eberline Teletector will not saturate, up to 30,000 R/hr (which is larger than the
maximum range).
It is easy to damage the instrument by bending the extendible tube.
EBERLINE RO-7
The RO-7 series instrument provides remote monitoring in high range beta and gamma
radiation fields. The RO-7 consists of a basic digital readout instrument, three
interchangeable detectors, and various interconnecting devices. The detectors may be
interconnected to the instrument by flexible cables of different lengths, by rigid
extensions of different lengths or by use of an underwater housing.
Detector Types
All three detectors are air-vented ion chambers contained in a plastic-lined (phenolic)
aluminum housing. The detector fill gas is air. The detector housing also contains other
electronics, such as an operational amplifier and detector identification circuitry.
The three available detectors are as follows:
• The RO-7-LD is a low-range, gamma-only detector with an active volume of
about 50 cm3 and dimensions of 2.5 cm diameter and 10.2 cm long.
DOE-HDBK-1122-99
Module 2.16 Radiation Survey Instrumentation Study Guide
2.16-18
• The RO-7-BM is a mid-range, beta/gamma detector, with beta window, that has
an active volume of about 7 cm3 and dimensions of 2.5 cm diameter and 5 cm
long.
• The RO-7-BH is a high-range, beta/gamma detector, with beta window, that has
an active volume of about 7 cm3 and dimensions of 2.5 cm diameter and 5 cm
long.
Each detector is labeled at the connector end of the detector.
• NOTE: Two small screws on the label are marked ZERO and CAL. These
should only be adjusted at calibration and must not be adjusted by the operator.
The RO-7 instrument is operated in the current mode of operation.
Instrument Operating Ranges
The operating range of the instrument is dependent on the detector that is connected to the
instrument.
• The range of the RO-7-LD detector is 0-2 R/hr.
• The range of the RO-7-BM detector is 0-200 R/hr.
• The range of the RO-7-BH detector is 0-20 kR/hr (20,000 R/hr).
Detector Shielding
All three detectors have a phenolic liner and aluminum housing.
Detector Window
The RO-7-BM and RO-7-BH detectors each have a 7 mg/cm2 Mylar window. The Lucite
cap for the beta window is 100 mg/cm2.
Types of Radiation Detected/Measured
As previously mentioned, the RO-7-LD detector measures only gamma and X-ray
radiation. Both beta and alpha radiation are shielded by the detector housing. The
neutron radiation response is insignificant due to the small size of the detector. The
actual detectors in the RO-7-BM and RO-7-BH detector assemblies are identical. Both
detect and measure gamma, X-ray and beta radiation. Alpha response is eliminated by
the 7 mg/cm2 window (same density thickness as the outer layer of skin). Neutron
radiation response is even smaller than the RO-7-LD due to the smaller detector volume.
Section 10
DOE-HDBK-1122-99
Module 2.16 Radiation Survey Instrumentation Study Guide
2.16-19
The energy response for the three detectors is as follows:
The RO-7-LD responds to photon radiation between 50 keV and 1.3 MeV (+20%). The
RO-7-BM and RO-7-BH detectors respond to photon radiation differently depending on
orientation and whether the Lucite cover is in place.
• Lucite cover off - 10 keV to 1.3 MeV (+20%)
• Lucite over on - 25 keV to 1.3 MeV (+20%)
• Shield on, from the side - 50 keV to 1.3 MeV (+20%)
• The beta response for the RO-7-BM and RO-7-BH detectors is for beta energies
>70 keV.
Operator-Adjustable Controls
The ON/OFF switch is the only range control because the instrument identifies the
detector model and adjusts the readout accordingly. A low battery condition is indicated
by a "colon" under the battery mark on the meter. The ZERO knob will zero the LCD
readout. A meter face light is turned on/off by the small switch in front of the pistol grip.
Detector Effective Center
No markings are provided for the detector effective center.
Specific Limitations/Characteristics
The response time of the basic instrument is 2.5 seconds to 90% of the final reading. The
correction factor for the true beta measurement is 1.5 as recommended by the
manufacturer.
Since the detector is air-vented, atmospheric temperature and pressure changes affect the
instrument reading. The instrument response will remain within +10% for the
temperature range of -20b to 160 bF. A correction table is available in the technical
manual for pressure changes. The detectors are air-vented but do not have a desiccant
pack. The detector should be kept dry and out of high humidity environments to prevent
leakage currents. Each detector has associated electronics designed for that particular
range. Overranging a detector may cause damage to the detector electronics. If the
instrument is not calibrated with the underwater housing and the housing is used, the
response will be about 5% low. The instrument reading should be multiplied by 1.05 to
obtain the corrected response.
DOE-HDBK-1122-99
Module 2.16 Radiation Survey Instrumentation Study Guide
2.16-20
2.16.04 Identify the following features and specifications for neutron detection and
measurement instruments used at your facility:
a. Detector type
b. Instrument operating range
c. Types of radiation detected/measured
d. Energy response
e. Operator-adjustable controls
f. Specific limitations/characteristics
Interconnecting devices from the detector to the instrument that are available from the
manufacturer are the:
• 15 ft flexible cable
• 60 ft flexible cable
• 2 ft rigid extension
• 5 ft rigid extension
• Stainless-steel underwater housing with 60 ft of cable.
NEUTRON DETECTORS
EBERLINE NRD NEUTRON SPHERE
The Eberline Neutron Rem Detector (NRD) sphere is a portable instrument for the
detection and measurement of the dose equivalent rate from neutron radiation.
Detector Types
The detector is the Eberline NRD (Neutron Rem Detector) sphere, which may be
connected to a PNR-4 (Portable Neutron Rem) or to an ESP (Eberline Smart Portable)
instrument by a coaxial cable. The NRD sphere is a 9 inch diameter, cadmium loaded,
polyethylene sphere with a BF3 proportional tube in the center of the sphere. The BF3
(boron trifluoride) detector design allows the detection of neutrons only, and the rejection
of other radiation. The thermal neutron capture reaction with the 10B results in gas
ionization pulses caused by the alpha particle from the reaction:
Section 11
1
0n + 10
5B S 73Li + 42He
or
10
5B (n,α) 7
3Li
DOE-HDBK-1122-99
Module 2.16 Radiation Survey Instrumentation Study Guide
2.16-21
The 9 inch diameter polyethylene sphere is used to moderate the neutrons. The
polyethylene has a high percentage of hydrogen which thermalizes the fast and
intermediate energy neutrons. Those neutrons that are thermalized in the sphere can be
detected in the BF3 tube. The cadmium loading is a thin sheet of cadmium placed at a
radius of about 7 cm inside the polyethylene sphere to help reduce the over response to
lower energy neutrons.
Instrument Operating Ranges
When the NRD is connected to the PNR-4 readout, there are two needles. One needle
covers the range up to 50 mrem/hr; above 50 mrem/hr, the second needle takes over and
covers the range up to 5000 mrem/hr. The NRD may also be connected to an Eberline
Smart Portable, ESP, which is discussed in lesson 2.17.
Types of Radiation Detected/Measured
Neutrons are measured. Alpha and beta radiation are not detected because they do not
penetrate the detector shielding. Gamma radiation passes through the detector shielding
but is rejected by the BF3 proportional chamber up to 500 R/hr (dependent on high
voltage setting and desired rejection level).
Since the BF3 detector is operated in the proportional region, the pulses from the alpha
particles are larger than pulses from other interactions and trigger a pulse height
discriminator in the instrument circuitry. The mode of operation for the instrument is the
pulse mode so that individual pulses can be discriminated and counted. Pulses from
gammas are rejected. Alpha and beta particles do not penetrate the polyethylene sphere.
Energy Response
The energy response curve for the instrument shows the relative response of the
instrument as compared to the theoretical dose equivalent. This instrument over-responds
to intermediate energy neutrons, and under-responds to relativistic neutrons.
Operator-Adjustable Controls
The only operator-adjustable control on the PNR4 is the OFF/ON/BAT switch which
turns the instrument on and off and allows a check of the battery.
No markings are provided for the detector effective center as the active volume of the
detector is centered in the sphere.
Specific Characteristics and Limitations
The response time depends on which decade of the scale is appropriate.
DOE-HDBK-1122-99
Module 2.16 Radiation Survey Instrumentation Study Guide
2.16-22
Response Times
First Decade 12 sec
Second Decade 6 sec
Third Decade 1.5 sec
Fourth Decade 0.3 sec
The detector is a sealed pressurized cylinder and is not affected by changes in humidity,
radioactive gases or changes in atmospheric density.
EBERLINE ASP-1 WITH NRD SPHERE
The Eberline ASP-1 with the NRD sphere is a microcomputer-based, analog-display,
portable neutron radiation survey instrument. The detector (Eberline NRD sphere) is
identical to the detector used with the PNR-4 readout packages. The mode of operation is
the pulse mode.
Instrument Operating Ranges
The overall range is 0-100 rem/hr and has a useable range of 1 mrem/hr - 60 rem/hr.
ASP-1 Ranges
0-1 mrem/hr
0-10 mrem/hr
0-100 mrem/hr
0-1,000 mrem/hr
0-10,000 mrem/hr
0-100,000 mrem/hr
Detector shielding is the same as previously mentioned for the NRD sphere. The energy
response is the same as previously mentioned for the NRD sphere.
Operator-Adjustable Controls
The OFF/BAT/HV/range switch has the following functions:
Section 12
• The OFF position turns the instrument off.
DOE-HDBK-1122-99
Module 2.16 Radiation Survey Instrumentation Study Guide
2.16-23
• The BAT position checks the instrument battery power supply.
• The HV position checks the applied high voltage to the detector and should match
the value listed on the special label on the instrument case.
The range markings are X1, X10, X100, X1K, X10K, and X100K with a meter scale of 0-
1.0.
The INTEGRATE/FAST/SLOW switch is a three position toggle switch with the
following functions:
• In the INTEGRATE position, the instrument will show the total dose equivalent
accumulated since the last time the instrument was reset to zero or turned off.
• In the FAST position, the response time selected by the microcomputer is for
typical survey work.
• In the SLOW position, the response is slower but with greater accuracy than the
FAST position.
The LIGHT/RESET switch is a three-position, spring-loaded toggle switch with the
following functions:
• The LIGHT position illuminates the meter face
• The RESET position will zero the meter reading for the current
mode (INTEGRATE/FAST/SLOW) setting of the instrument.
The RESET switch will cause the "standard current" value to be displayed if held for
5 seconds while in the FAST or SLOW mode. The SPEAKER is a two-position toggle
switch for turning the external speaker on and off. Acoustic (airline-type) head phones
can be plugged into the speaker cover on the top of the instrument. As previously
mentioned, the NRD sphere has no effective center markings.
Specific Limitations/Characteristics
The response time of the instrument is controlled by the microcomputer and is based on
the input count rate and whether the mode switch is in FAST or SLOW. In the FAST
position, the instrument response time varies between one and ten seconds. In the SLOW
position, the instrument response time will vary up to a maximum of 29 seconds. No
correction factors are required to correct the displayed value. The sealed detector is not
affected by changes in atmospheric density, humidity or radioactive gases. The
instrument has a microcomputer controlled "overrange" indication. When the radiation
rate exceeds the useful range of the detector, the computer will cause an overrange alarm.
DOE-HDBK-1122-99
Module 2.16 Radiation Survey Instrumentation Study Guide
2.16-24
When the instrument alarms, the meter needle will sweep back and forth and an
interrupted tone will sweep in the speaker.
SUMMARY
This lesson has covered the specifications, features and limitations for the portable
radiation survey instruments that may frequently be used by the RCT. This knowledge
should be used to properly select and operate the instruments to ensure that the data
obtained is accurate and appropriate The appropriate, accurate data is then used to
properly assign external exposure controls.
Module Number: 2.16